| brand | Other brands | Price range | 10000 to 30000 |
|---|---|---|---|
| Instrument type | laboratory | Flow range | large flow |
| Origin category | domestic |
1Y09-310 LCD AC laser dust particle counteroverview
This product is a dust particle counter jointly developed by our factory and Nanjing University of Science and Technology. The control method has been upgraded from old-fashioned button based control to touch screen control, with an 8-inch color LCD touch screen (display size 162mm wide * 121mm high, display ratio 4:3), and a large visual interface making operation more convenient and intuitive. The product adopts imported German Thomas air pump, which has lower noise and less electromagnetic interference; The exterior design is atmospheric and simple, with a small volume and light weight,Easy to use, with a fully stainless shell that better meets GMP requirements.
IIY09-310 LCD AC laser dust particle counterTechnical indicators:
model |
Y09-310 LCDLaser dust particle counter |
sample flow |
28.3L/min ±5% |
Particle size grading |
0.3um,0.5um,1um,3um,5um,10um |
laser light source |
Long life laser diode with a lifespan of 10 years |
Imported air pump |
Continuous and stable operation, energy-saving and environmentally friendly |
Verification standards |
GB/T6167-2007,JJF1190-2008 |
Repetitive relative error |
<±10% |
Particle size distribution error |
<±30% |
Sampling concentration |
35000 pieces/liter |
cleanliness recovery characteristic |
≤ 10 minutes (counting to zero for 3 consecutive times within 10 minutes, 95% confidence level) |
sample time |
User self setting (5-1800 seconds) |
SAMPLE DELAY |
User self set (1-100s) interval 1-1800s |
UCL sampling frequency |
2∽50 |
UCL sampling points (position) |
2∽50 |
Data memory capacity |
1 to 5000 sets of measurement data, searchable |
count mode |
Total number of |
screen display |
Color 8-inch touch screen |
Printing method |
Built-in Thermal Printer |
power supply |
Communication 220V/50Hz |
Cleanliness level determination criteria |
ISO14644-1, GMP Static, GMP Dynamic |
Temperature and humidity |
Optional |
Data communication interface |
USB |
environment |
Usage environment: 0-40 ℃, relative humidity 10-90% |
|
environment Standard accessory |
Storage environment: -30 ℃ to 45 ℃, relative humidity 0 to 90% |
|
environment Standard accessory External dimensions |
Waiting for power sampling head, connecting hose (inner diameter 6mm) with self-cleaning filter, tripod, printing paper, portable protective box, power cord, USB flash drive, USB data cable, user manual, original factory calibration report, certificate of conformity |
Length 315 X Width 260 X Height 230 | |
material |
SUS304 stainless steel |
Weight (excluding packaging) |
7Kg |
IIIUsage conditions:
1. Clean room environment with no strong magnetic field interference around; Do not work in the atmosphere for a long time to prevent sensor contamination
2. Power supply: AC220V50Hz。
4after-sale service
One year warranty and free software upgrade