Single flame type (reserved graphite furnace interface) atomic absorption spectrophotometerIntroduction:A precision instrument mainly used to detect the content of trace metal elements in various substances. It can detect more than 40 elements including manganese, iron, copper, zinc, lithium, sodium, magnesium, iron, potassium, calcium, chromium, cobalt, nickel, arsenic, selenium, silver, cadmium, indium, tin, antimony, platinum, gold, lead, etc. It is mainly detected by flame method, and the detection limit can generally be PPM level (parts per million content)
ZCA-1000SF(Standard single flame) atomic absorption spectrophotometer
atomic absorption spectrophotometerIntroduction:A precision instrument mainly used to detect the content of trace metal elements in various substances. Can detect the following elements: manganese, iron, copper, zinc, lithium, sodium, magnesium, iron, potassium, calcium, chromium(gè)Cobalt, nickel, arsenic selenium(xī) Silver, cadmium, indium, tin, antimony, platinum, gold, lead class40Multiple, mainly detected by flame method, the detection threshold can generally bePPMGrade (parts per million content)
1、 Performance indicators
1Optical system
wavelength range190nm—900nm Automatic adjustment
Grating line density 1800strip
monochromator anaberrationCzerny-TurnerType monochromator
spectral bandwidth 0.1、0.2、0.4、1.0、2.0nmMulti level automatic switching
Wavelength accuracy±0.20nm
Wavelength Repeatability0.1nm
baseline drift static state0.005A/30min dynamics0.008 A/30min
detector Wide spectral range photomultiplier tube
resolution Can distinguish manganese lamps279.5nmand279.8nmDouble lines, and the energy of the valley between the two spectral lines should be less than20%
2Light source system
lamp base 3lamp base
Lamp power supply mode400Hzsquare-wave pulse
Lamp current adjustment range0 — 10mAaverage current
3Atomization system
Flame analysis method
detection limit (Cu) 0.006μg/ml
burner100mmAll titanium metal burner
precisionRSD≤1%
sprayer Metal sheathed high-efficiency glass spray
Atomization chamber Corrosion resistant all plastic atomization chamber
Security measures Equipped with multiple automatic safety protection functions, acetylene leak alarm and automatic shutdown system
Flameless(Graphite furnace)Analysis method
detection limit(Cd)≤4pg
4Flame analysis method and measurement method
air-Acetylene flame method, hydride generator atomic absorption method
Concentration calculation method Automatically fitting the working curve, automatically correcting sensitivity, and automatically calculating concentration and content.
Repeated measurement times1~99Next, automatically calculate the mean, content, standard deviation, and relative standard deviation.
Result printing Print test data or final analysis report. The results can be exported asExcelWorksheet.
software environmentWindows XP/VistaOperating system, professional Chinese software
2、 Instrument parameters
power supplyAC 220V/50HZ
power150VA
Net weight of the instrument55kg
Gross weight of the instrument80kg
Instrument volume1110mm(Horizontal)× 400mm(Depth)× 500mm(High)
Packaging volume1350mm(Horizontal)× 600mm(Depth)× 850mm(High)
Ambient Temperature10~35℃
Working environment humidity≤80%
3、 Standard configuration
1. Host
2Low noise oil-free air compressor
3Three hollow cathode lamps
4Operating software
4ZCA-1000Design features of standard atomic absorption spectrophotometer
1. Total reflection optical system, full band chromatic aberration reduction. After adjusting the focal point, any wavelength can be changed, and the focal point will no longer change. The wavelength will be automatically scanned and peak searched.
2. Featuring deuterium lamp buckle background and self-priming buckle background.
3. The uniquely designed all titanium burner has smooth combustion and fast thermal equilibrium speed.
4. Automatically adjust negative high voltage, lamp current, element lamp position to maximize energy, and automatically protect against gas leakage and flameout.3Support hollow cathode lamp holder. Automatic control system, capable of automatically scanning wavelengths.1Preheat one element lamp and preheat the other two element lamps.
5. Graphite furnace heating: Users can freely set the heating curve according to their needs.
|
Flame method Sensitivity and detection limit of commonly measured elements
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||