Total phosphorus measuring instrument
The German WTW total phosphorus measuring instrument adopts the molybdenum blue colorimetric method, which has two-point calibration, automatic digest
Product details
GermanyWTWTotal phosphorus measuring instrumentTresCon TP
For applications that require accurate testing and real-time adjustment of water treatment processes,TresConIt was their wise choice. Compared to on-site direct input sensors,TresConIt has the characteristics of more accurate and stable testing, and stronger anti-interference.TresConAdopting modular design, * allowing for more3The arbitrary combination of modules and multi parameter monitoring system have unparalleled cost advantages.
Testing Principles
The total phosphorus module is divided into two parts, the * * part is a high-temperature chemical digestion module used to heat and digest samples, and the * * part is an analysis module used to test phosphate concentration.
During the digestion process, all valence states of phosphorus compounds are converted into orthophosphate that can be analyzed by photometry. The reaction is based on the oxidation of other valence states of phosphorus to orthophosphate by persulfate under acidic conditionsPO43-High temperature and high pressure conditions can greatly shorten the digestion time.
The molybdenum blue colorimetric method is used to test the concentration of orthophosphate: molybdate is added to the sample, and a chemical reaction is carried out to make the sample blue. The color change is sensed by a photometer and then converted into concentration. The entire process is carried out in an orderly manner under the control of the controller, thus ensuring high operational safety.
Total phosphorus measuring instrumentTresCon TPTechnical Parameter
| |
Sample pretreatment |
A separate sample preprocessor is required, such asPurconperhapsPF105
|
Sample supply |
Equipped with a standard flow channel, the overflow sample volume is sufficient3An analysis module can be used, and it can also be installed simultaneously3A circulation channel is used for testing3A different testing point
|
Outputs |
3Group isolation0/4-20 mAOutput,12Passive relay output,RS232andRS485
|
power supply |
230 VAC ± 10%, 50 Hz / 115 VAC ± 10%, 50 – 60 Hz |
Surrounding environment |
Storage temperature:25 . .. 60 °CWorking temperature:0 .. 40 °C
|
authentication |
CE, DIN-GOST |
Protection level |
IP54,IEC 1010-1/EN 61010-1, Class 1
|
weight |
27 kg(Excluding analysis module)Each analysis module is reconfigured10 kg,Installation bracket weight approximately25 kg
|
TresConAutomation function
| |
AutoClean |
Automatic cleaning function, tested the sewage plant discharge outlet without any problems |
AutoCal |
Automatic calibration function, adjustable calibration interval, ensuring better testing accuracy |
AutoKorr |
Automatic background value correction function to improve the accuracy of photometric analysis |
AutoFlow |
Automatic container liquid level monitoring function, automatically prompts for filling reagents or pipeline blockage |
AuthTherm |
Automatic constant temperature function |
Test range |
Total phosphorus:0.01 …3.00 mg/l, 0.3 … 100 µmol/l 0.01 … 6.00 mg/l, 0.3 …200 µmol/l (1:1dilute) |
resolution |
Test range:0.01 …3.00 mg/l: 0.01 mg/l 0.30 …100 µmol/l: 0.1 µmol/l |
accuracy |
±3%testValue ±0.05 mg/l P
|
measuring principle |
Reference beam colorimetric method after digestion |
test method |
Molybdenum blue colorimetric method |
Method deviation coefficient |
1.5% |
reaction time |
<4minute
|
Test Interval |
10,15,20,25,30,60Minutes optional; DINDissolving method(120 °C, thirtyMinutes)
|
correct |
automatic2Point Correction
|
Sample consumption and requirements |
10 ml/order,Suspended particles<50 mg/l
|
Reagent consumption |
Reaction reagents(A,B,C,D)Available60Every day (every60Measured every minute),
Standard solution and cleaning solution are available60order
reagentA、B、C、D:10/15/20/30/60Oh my god, every10/15/20/30/60Minute measurement interval
Standard liquid1.5 l:70Oh my god,24Hourly calibration interval
cleaning solution1.5 l:60Oh my god,24Hourly cleaning interval
|
service interval |
every3Month
|
Online inquiry