1TX-YBS-DT precision digital pressure gaugesummary:
TX-YBS-DT precision digital pressure gauge is the latest portable high-precision and multifunctional pressure calibration instrument launched by our company. It can be used as a high-precision measuring instrument, calibration instrument to calibrate and verify pressure/differential pressure transmitters, pressure sensors, ordinary/precision pressure gauges, etc. It also has the function of potentiometers, precision millivoltmeters, precision ammeters, cold junction thermometers and thermocouple calibration tables. It can be widely used in industrial and mining sectors and research institutions such as petroleum, chemical, metallurgical, power, metering, automation, etc.
This instrument measures pressure through a high-precision intelligent pressure module; It also has a six and a half digit digital multimeter and signal source with DC input and output functions, including measurement and output of voltage and current in different ranges.
One Main functions:
1. Pressure measurement.
2. Voltage measurement.
3. Voltage output.
4. Current measurement.
5. Current output.
Precision temperature measurement and simulation of eight commonly used thermocouples, including B, E, J, K, N, R, S, T, etc.
7. Thermocouple calibration table.
twoTX-YBS-DT precision digital pressure gaugeTechnical indicators:
1. Pressure:
The pressure module of Tianxin Instrument is divided into gauge pressure, differential pressure, absolute pressure and other types, with the main range being:
(1) Gauge pressure: -100KPa to 100MPa (various levels)
(2) Differential pressure: -100KPa~1MPa (various levels)
(3) Absolute pressure: 0~110KPa (various levels)
Main pressure range measurement accuracy: 0.05% F · S; 0.02%F·S
2. Voltage:
Measurement: The accuracy of each range within 0-30V is ± (0.005% RD+0.005% FS) for the 0-5V range,
Input impedance: › 5M Ω.
Output: 0-15V, accuracy: ± (0.005% RD+0.005% FS),
Output impedance: ‹ 1 Ω
3. Current:
Measurement: 0-30mA, accuracy: ± (0.005% RD+0.005% FS),
Input impedance: 5 Ω (with additional fuse resistance).
Output: 0-30mA, accuracy: ± (0.005% RD+0.005% FS).
4. Thermocouple temperature indication
|
model
|
Measurement range (℃)
|
Indication accuracy% FS (guaranteed temperature range ℃)
|
|
K
|
-270~1360
|
±(0.005% RD+0.005% FS)(-150~1360)
|
|
N
|
-270~1280
|
±(0.005% RD+0.005% FS) (-100~1280)
|
|
R
|
-50~1760
|
±(0.005% RD+0.005% FS) (50~1760)
|
|
S
|
-50~1760
|
±(0.005% RD+0.005% FS) (50~1760)
|
|
T
|
-270~400
|
±(0.005% RD+0.005% FS) (-100~400)
|
|
B
|
50~1800
|
±(0.005% RD+0.005% FS) (500~1800)
|
|
E
|
-170~1000
|
±(0.005% RD+0.005% FS) (-150~1000)
|
|
J
|
-210~1200
|
±(0.005% RD+0.005% FS) (-150~1200)
|
Internal rechargeable battery 6V, power consumption: ‹ 0.6W.
External power supply: 220V, 50Hz.
6. Working environment:
Temperature: 5-50 ℃, humidity: ‹ 90%.
7. Size: 250mm × 250mm × 100mm.