Shanghai polyol determination gas chromatograph manufacturer
The principle of gas chromatography analysis:
Qualitative analysis by chromatography: Is there a standard substance for the analyzed component, known or unknown?
Chromatographic quantitative analysis: in that range - constant (10-1 `~10-3); Semi trace (10-3~10-5); Trace (10-5-10-7); Trace (10-6~10-9) or ultra trace (≤ 10-9)
The accuracy of chromatographic quantification and analysis is much simpler if semi quantitative requirements are met.
Shanghai polyol determination gas chromatograph manufacturerinstrumentMain technical indicators:
● Interface display: 5.7-inch industrial color LCD screen
Temperature control area: 8 channels, can control dual column boxes, each column box has an independent rear door design.
Temperature control range: 4 ℃ to 450 ℃ above room temperature, increment: 1 ℃, accuracy: ± 0.01 ℃
● Program heating order: 16 orders, 17 platforms
● Lift rate: 0.1-39 ℃/min (ordinary type); 0.1-80 ℃/min (high-speed type)
External events: 8 channels; The auxiliary control output consists of 2 channels
● Sample injector types: packed column injection, capillary injection, gas valve injection, headspace injection, thermal analysis injection, blowdown collector injection, flash evaporation injection, etc
● Number of detectors: up to 9 high-performance detector options are available, including FID, TCD, ECD, FPD, NPD, PID, PDI ID, ZD, DID, etc., and 3 types of detectors can be installed simultaneously
● Gas path control: Mechanical valve control mode and EPC mode can be selected, or the two modes can be mixed and used.
EPC and EFC working modes: 2 types; Constant current mode, constant voltage mode
EPC and EFC working gases: 5 types; Nitrogen, hydrogen, air, helium, argon
EPC and EFC lift: 4th order
EPC and EFC control range: pressure: 0-0.6MPa; Flow rate 0-100mL/min or 0-500mL/min (air)
● Pressure sensor:
Accuracy:<± 2% of full scale
Reproducibility:<± 0.05 KPa
Temperature coefficient:<± 0.01 KPa/° C
● Flow sensor:
Accuracy:<± 5% of full scale
Reproducibility:<± 0.5% (full scale)
Range: 0-500mL/min
● Start injection: manual or automatic selection
● Communication interface: Ethernet: IEEE802.3
Power supply: 220V ± 10%, 50Hz; 2500W (maximum)
● Volume: 572 × 552 × 465 (height) mm
Weight: 50kg (approximately)
Technical specifications of detector:
Hydrogen flame ionization detector (FID)
★ Detection limit: Mt ≤ 3 × 10-12g/s (n-hexadecane-isooctane solution);
★ Baseline noise: ≤ 5 × 10-14A
★ Baseline drift: ≤ 1 × 10-13A/30min
★ Linear range: ≥ 106
Thermal conductivity detector (TCD):
★ Sensitivity: S ≥ 3500mV • ml/mg (benzene solution) (standard), maximum sensitivity 10000mV • ml/mg (equipped with electronic amplifier board, 2, 4, 8 times optional)
★ Baseline noise: ≤ 10 μ V
★ Baseline drift: ≤ 30 μ V/30min
★ Linear range: ≥ 104
Electron Capture Detector (ECD):
★ Detection limit: ≤ 1 × 10-14g/s
★ Baseline noise: ≤ 3 × 10-13A
★ Baseline drift: ≤ 0.2mV/30min
★ Linear range: ≥ 103
★ Radioactive source: 63Ni
Flame photometric detector (FPD):
★ Detection limit: (S) ≤ 5 × 10-11g/s, (P) ≤ 1 × 10-12g/s;
★ Baseline noise: ≤ 3 × 10-13A
★ Baseline drift: ≤ 0.2mV/30min
★ Linear range: ≥ 103 (S), 102 (P)
Helium ionization detector (PDIHID):
★Discharge mode: Pulse discharge
★Baseline noise: ≤ 0.1mv
★Baseline drift: ≤ 0.5mv/30min
★Detection limit: ≤ 1.0 × 10-10g/ml
Photoionization detector (PID)
★ Photon energy 10.6eV
★ Light source lifespan ≥ 3000h
★Detection limit: ≤ 1ug/m3 (benzene)
★ Baseline noise: ≤ 8 × 10-14A
★Linear range:≥105
Analysis conditions:
Chromatography column: 30 * 0.53 * 0.5 PEG
Column temperature: 70℃→10min→10℃/min→200℃→20min
Carrier gas: N2 30min/sec
Sample injector: 250℃split injection
Detector: FID 250℃
Sample analysis: polyol mixture