Su Rui Automation Instrument (Jiangsu) Co., Ltd
Home>Products>Sewage pipeline flow meter
Sewage pipeline flow meter
Electromagnetic flow meters are suitable for measuring and controlling process flow in industrial and agricultural production companies such as chemic
Product details

详情页主图.jpg

产品简介.jpg

The sewage pipeline flow meter consists of two parts: sensors and converters. It works based on Faraday's law of electromagnetic induction and is used to measure the volumetric flow rate of conductive liquids with a conductivity greater than 5 μ S/cm. It is an inductive instrument for measuring the volumetric flow rate of conductive media. In addition to measuring the volumetric flow rate of general conductive liquids, it can also be used to measure the volumetric flow rate of strongly corrosive liquids such as strong acids and alkalis, as well as homogeneous liquid-solid two-phase suspensions such as mud, slurry, and pulp. Widely used in industrial sectors such as petroleum, chemical, metallurgical, textile, papermaking, environmental protection, food, as well as municipal management, water conservancy construction, river dredging and other fields for flow measurement.

111.jpg工作原理.jpg

The measurement principle of sewage pipe network flowmeter is based on Faraday's law of electromagnetic induction. The sensor part is composed of coils and electricity4e4a20a4462309f7e52321fc7c0e0cf3d6cad694.jpgComposed of polar and insulating liningWhen measuring, the excitation coil in the sensor is energized to generate a magnetic field. When the conductive fluid passes through the magnetic field, a small induced electromotive force is generated due to the cutting force of the magnetic field lines. These small induced electromotive forces are collected by the electrodes and transmitted to the converter part of the instrument for signal amplification, correction and other operations. Then, they are converted intoThe corresponding flow data is ultimately displayed on the instrument.

When a conductive fluid flows through a magnetic field perpendicular to the flow direction, the conductive liquid induces an induced voltage E proportional to the average flow velocity. The induced voltage is detected by two electrodes in direct contact with the fluid, amplified, filtered, and shaped by a converter, and sent to the MCU to display and output control the instantaneous flow rate and cumulative flow rate.

E=KBVD In the formula: E - Induced voltage K - Instrument constant B - Magnetic induction intensity V - Average flow velocity inside the measuring tube D - Diameter of the flowmeter

产品特点.jpg

The instrument structure is simple and reliable, with no movable parts and a long working life

No flow blocking components, no pressure loss or fluid blockage.

No mechanical inertia, fast response, good stability, can be applied to automatic detection and adjustment of programmable control systems.

The measurement accuracy is not affected by the type of measured medium and its physical parameters such as temperature, viscosity, pressure, etc.

Different combinations of PTFE or rubber lining materials and electrode materials such as HC, HB, 316L, Ti can adapt to the needs of different media.

All digital processing, strong anti-interference ability, reliable measurement, high accuracy, and a flow measurement range of up to 150:1.

Full Chinese character menu operation, easy to use and simple to operate.

High definition backlit LCD display.

It has bidirectional flow measurement and bidirectional total accumulation function. It has three internal integrators that can display forward accumulation value, reverse accumulation value, and difference accumulation value respectively.

Output mode: bidirectional output function of current and frequency, and RS-485, MODBUS, HART interfaces.

Using SMD devices and surface mount technology, the circuit has high reliability.

技术参数.jpg

Technical data of sewage pipe network flowmeter and sensors

execution standard

JB/T9248- l999

Nominal Diameter

10. 15, 20, 25, 32, 40, 50, 65, 80, 100, 125, 150, 200, 250, 300, 350, 400, 450, 500, 600, 700, 800, 900 l000、l200、1400、1600、 l800

flow velocity

15m/s

accuracy

Level 0.5, Level 1.0 (differentiated by caliber)

Fluid conductivity

≥5uS/cm

ambient temperature

Sensor: (-40~+80) ℃, Converter: (- l5~+50) ℃

Lining material

Polytetrafluoroethylene, chloroprene rubber, polyurethane, perfluoroethylene propylene (F46)

fluid temperature

≤120℃

electrode material

316L、 Hastelloy C, Hastelloy B, titanium, tantalum, platinum/iridium alloy, stainless steel coated with tungsten carbide

flange material

Carbon steel, stainless steel

Shell protection

IP65

Explosion-proof marking

ExⅡBT6 Gb

Technical data of flow meter converter for sewage pipeline network

power supply direct current (18~30)V
communication (85~265)V (45~63)Hz
power <20W (matched with sensor)
Internal integrator Positive flow, reverse flow, and differential flow all have total integrators




output signal

(Programmable)

current output

Output signal: Two way bidirectional, fully isolated (0~10) mA/(4~20) mA

Load resistance: (0-10) mA, (0-1.5) K Ω; (0-20) mA, (0-750) Ω

Basic error: Add ± 10 μ A to the basic measurement error mentioned above

Frequency output

Forward and reverse flow output, with an upper limit of output frequency that can be set within 1-5000Hz

● Transistor collector open circuit bidirectional output with photoelectric isolation

The external power supply should not exceed 35V, and the collector current should be 250mA when conducting

pulse output

Forward and reverse flow output, with an upper frequency limit of up to 500cp/s

● Automatic pulse width setting or 20ms square wave

● Transistor collector open circuit bidirectional output with photoelectric isolation

The external power supply should not exceed 35V. When conducting, the collector current should be 250mA

Flow direction indication output Can measure fluid flow in both directions and determine the direction of fluid flow
alarm output

Two transistor collector open circuit alarm outputs with photoelectric isolation

The external power supply should not exceed 35V, and the collector current should be 250mA when conducting

● Alarm status: Fluid empty pipe, excitation disconnection, flow exceeding limit


Communication interface

RS-485、 MODBUS、 HART, Equipped with lightning protection
Damping time Within (1~64) s
operational condition Environmental temperature: (-10~60) ℃, relative humidity: 5%~90%

Flow measurement range of sewage pipe network flowmeter:

Caliber (mm) Measurement range (m 3/h) Caliber (mm) Measurement range (m 3/h) Caliber (mm) Measurement range (m 3/h)
DN10 0.14 ~ 1.4 DN125 22.08 ~ 441.56 DN700 692.37 ~ 13847.40
DN15 0.32 ~ 6.36 DN150 31.79 ~ 635.85 DN800 904.32 ~ 18086.40
DN20 0.57 ~ 11.30 DN200 56.52 ~ 1130.4 DN900 1144.53 ~ 22890.60
DN25 0.88 ~ 17.66 DN250 88.31 ~ 1766.25 DN1000 1413.00 ~ 28260.00
DN32 1.45 ~ 28.94 DN300 127.17 ~ 2543.40 DN1200 2034.72 ~ 40694.40
DN40 2.26 ~ 45.22 DN350 173.09 ~ 3461.85 DN1400 2769.48 ~ 55389.60
DN50 3.53 ~ 70.65 DN400 226.08 ~ 4521.60 DN1600 3617.28 ~ 72345.60
DN65 5.97 ~ 119.40 DN450 286.31 ~ 5722.65 DN1800 4578.12 ~ 91562.40
DN80 9.04 ~ 180.86 DN500 353.25 ~ 7065.00

DN100 14.13 ~ 282.6 DN600 508.68 ~ 10173.6

衬里选择.jpg

未标题-1.jpg


电极选择.jpg

电极材料.jpg

安装工艺.jpg

安装1.jpg安装2.jpg

Requirements for external environment of sewage pipeline flow meter:

  1. Flow meters should be avoided from being installed in places with large temperature changes or high temperature radiation from equipment. If installation is necessary, insulation and ventilation measures must be taken.

  2. The flowmeter should be installed indoors. If it must be installed outdoors, it should be protected from rainwater, waterlogging, and direct sunlight. Moisture and sun protection measures should be taken.

  3. Flow meters should be avoided from being installed in environments containing corrosive gases, and ventilation measures must be taken when installation is necessary.

  4. For the convenience of installation, maintenance, and upkeep, there should be ample installation space around the flowmeter.

  5. The installation site of the flowmeter should avoid strong magnetic fields and vibration sources. If the pipeline vibration is large, there should be fixed pipe supports on both sides of the flowmeter.

Requirements for straight pipe section of sewage pipe network flowmeter:

In order to improve the effects of eddy currents and flow field distortions, there are certain requirements for the length of the straight pipe sections before and after the installation of the flowmeter, otherwise it will affect the measurement accuracy (rectifiers can also be installed, avoiding installation near the regulating valve and half open valve as much as possible).

Requirements for process pipes: The flowmeter has certain requirements for the upstream and downstream process pipes of the installation point, otherwise it will affect the measurement accuracy.

The inner diameter of the upstream and downstream process pipes should be the same as that of the sensor, and should meet the requirement of 0.98DN ≤ D ≤ 1.05DN (where DN is the inner diameter of the sensor and D is the inner diameter of the process pipe). The process pipe and sensor must be concentric, with a coaxial deviation of no more than 0.05DN. The requirements for the bypass pipe are: to facilitate the maintenance of the flow meter, a bypass pipe must be installed for the flow rate. In addition, for heavily polluted fluids and flow meters that need to be cleaned and cannot be stopped, a bypass pipe must be installed.

Installation requirements for flow meters on pipelines:


Sensor grounding: In order to ensure reliable operation of the instrument, improve measurement accuracy, and avoid interference from external parasitic potentials, the sensor should have a good separate grounding wire with a grounding resistance of less than 10 Ω. If there is insulation layer or non-metallic pipeline in the pipeline connecting the sensor, grounding rings or built-in grounding electrodes should be installed on both sides of the sensor.

Grounding method on metal pipelines: If there is no insulation layer inside the metal pipeline, ground it according to the following diagram.



Grounding method on plastic pipes or insulated or painted pipes: Grounding rings or built-in grounding electrodes should be installed on both ends of the sensor to short-circuit the measured medium flowing inside the pipe to the ground and have zero potential. Otherwise, the electromagnetic flowmeter will not function properly.