RX6160 Combination Multi Channel Verification Instrument
The combined multi-channel calibrator RX-6160 is an open and innovative desktop calibration instrument
Product details
The combined multi-channel calibrator RX-6160 is an open and innovative desktop calibration instrument. It adopts a working mode of 1 host+signal module (up to 8 channels), where each signal is completely isolated and used simultaneously. Effectively achieving customer on-demand selection, avoiding waste, and expanding functionality at any time, it can maximize the satisfaction of customer needs. Multiple communication methods (supporting Rs232, Lan, USB, GPIB) can meet customers' various application needs.
The launch of this product effectively solves the problem of fixed and limited functions of calibration instruments in the current market. Users can freely combine different signal modules according to their needs to create a product that meets practical applications.
RX6160Host features:
1. Display screen: 4.3-inch color screen (resolution 480 * 272).
2. Number of supported signal modules: 8, fully isolated from each other.
3. Display mode: Single channel, 2-channel, 4-channel, 8-channel optional, supports cyclic display (interval time can be set).
4. Output mode: Supports three forms simultaneously: typing values, cursor shifting, and knob adjustment.
5. Support% display.
6. Support graphical display of historical data.
7. Supports displaying maximum, minimum, and average values.
8. Support measurement value locking.
9. Support step, step, ramp, fixed-point output and other methods.
10. Friendly human-computer interaction interface, convenient operation mode; The host software can be upgraded by the customer themselves.
11. Provide multiple interface functions to support customers' secondary development.
12. Data memory function: up to 10000 sets of data storage and memory functions.
13. Support logical conditional response between modules, response speed: 100ms, response speed with external I/O ports: 20 µ s.
14. Communication method: RS232, LAN, USB are standard configurations, GPIB can be optionally configured, and communication protocols are provided for free.
15. Power supply mode: 220 V.AC ± 10%, 50HZ, optional 100/110V. AC ± 10% (specified when ordering).
16. Host power consumption: 5W for a single machine, 10W when using USB function.
17. Maximum allowable power consumption of signal module: A single channel module can consume up to 10W, and a total of 8 signal modules can consume up to 20W.
Technical specifications of signal module:(All ports support 220V AC misconnection, 36V DC misconnection, and short-circuit self recovery function)
| Module Name | model | function | Range of measurement (resolution) | Accuracy (reading%)+Range%) | Maximum power consumption |
| Current measurement output module | 60P01 | current output | 0~30 mA(0.0001 mA) | 0.05+0.005 | 2W |
| Current Measurement | -30mA~30 mA(0.0001 mA) | ||||
| Current measurement output module | 60P02 | current output | Same as "30P01" | 0.02+0.005 | 2W |
| Current Measurement | |||||
| Voltage and resistance measurement output module | 60P03 | voltage measurement | -200mV~200mV(0.001mV) -2V~2V(0.00001V) -10V~10V(0.0001V) -100V~100V(0.001V) |
0.05+0.005 | 1.5W |
| Voltage Output | -15mV~200mV(0.001mV) 0V~11V(0.0001V) |
||||
| resistance measurement | 0~450Ω(0.001Ω) 420Ω~4500Ω(0.01Ω) |
||||
| Resistance output | 0~450Ω(0.001Ω) 420Ω~4500Ω(0.01Ω) |
0.05+0.01 | |||
| Voltage and resistance measurement output module | 60P04 | voltage measurement | Same as "30P03" | 0.02+0.005 |
1.5W |
| Voltage Output | |||||
| resistance measurement | |||||
| Resistance output | 0.02+0.01 | ||||
| Frequency module | 60P05 | frequency | 1~99.9999Hz(0.0001Hz) 100~999.999Hz(0.001Hz) 1k~9.99999kHz(0.01Hz) 10k~99.9999kHz(0.1Hz) |
0.01+0.005 | 0.5W |
| Frequency output | 0~200.000Hz(0.001Hz) 0~2000.00Hz(0.01Hz) 0~20.0000kHz(0.1Hz) |
0.01+0.005 | |||
| Multi signal output module | 60P06 | Voltage Output | -15mV~200mV(0.001mV) 0V~11V(0.0001V) |
0.05+0.005 | 2W |
| current output | 0~30 mA(0.0001 mA) | ||||
| Resistance output | 0~450Ω(0.001Ω) 420Ω~4500Ω(0.01Ω) |
0.05+0.01 | |||
| Frequency output | 0~200.000Hz(0.001Hz) 0~2000.00Hz(0.01Hz) 0~20.0000kHz(0.1Hz) |
0.01+0.005 | |||
| Multi signal measurement module | 60P07 | voltage measurement | -200mV~200mV(0.001mV) -2V~2V(0.00001V) -10V~10V(0.0001V) -100V~100V(0.001V) |
0.05+0.005 | 1W |
| Current Measurement | -30mA~30 mA(0.0001 mA) | ||||
| resistance measurement | 0~450Ω(0.001Ω) 420Ω~4500Ω(0.01Ω) |
||||
| frequency | 1~99.9999Hz(0.0001Hz) 100~999.999Hz(0.001Hz) 1k~9.99999kHz(0.01Hz) 10k~99.9999kHz(0.1Hz) |
0.01+0.005 | |||
| Thermal multi signal output module | 60P08 | Voltage Output | Same as "30P06", add thermocouples J, K, T, E, R, S, B, N, and temperature tables for thermal resistors Pt100, Pt1000, Cu50, and Cu100 | 0.05+0.005 | 2W |
| current output | |||||
| Resistance output | 0.05+0.01 | ||||
| Frequency output | 0.01+0.005 | ||||
| Thermal multi signal measurement module | 60P10 | voltage measurement | Same as "30P07", add thermocouples J, K, T, E, R, S, B, N, and temperature tables for thermal resistors Pt100, Pt1000, Cu50, and Cu100 | 0.05+0.005 | 1W |
| Current Measurement | |||||
| resistance measurement | |||||
| frequency | 0.01+0.005 | ||||
| i/o module | 60P12 | I/O signal | 8-channel switch input and output signals, response time<20 µ s | 0.5W | |
| Power output module | 60P13 | Power Output | Adjustable from 0 to 36V, maximum load current 200mA | 8W | |
- The above function accuracy is applicable within a temperature range of 18~38 ℃.
- The "current output" in the above functions supports both active and passive output modes, with a maximum load of 1k Ω for 20mA.
- The "current measurement" function mentioned above supports loop measurement, with a loop voltage of 24V ± 0.5V.
- The "resistance measurement" function mentioned above supports three wire and four wire methods, and supports on/off testing.
- The "frequency measurement" function mentioned above has a measurement sensitivity of 10Hz to 10kHz and Vp-p ≥ 1V; The remaining Vp-p is ≥ 2V; Waveform: Square wave, additional 5 counts are required for errors in other waveforms, which can directly measure the mains frequency.
- The above function includes "frequency output", with output waveforms such as square waves, sine waves, etc., and adjustable amplitudes from 0 to 10V.
- The "resistor output" function allows external excitation current of 0.15~4mA.
- The "voltage output" in the above functions all support capacitive loads.
- PT100 thermistor (model 30A01) is selected for temperature compensation, with a compensation accuracy of 0.2 ℃ and a compensation temperature range of -40 ℃ to 60 ℃.
- Attention to signal board selection: The cumulative total power of the signal board should be ≤ 20W.
- As the types of signal sampling samples may increase at any time, please follow our official website for the latest releases.
Customized services:
1. Provide customization of special signal modules;
2. Module pin definitions can be provided for users to develop their own signal modules;
3. Redefine panel buttons or display interfaces according to customer needs;
4. Develop upper level management software or multi machine logic condition setting function according to customer needs.
Types and Accuracy Table of Thermocouples and Thermistors
| Graduation | Range of measurement | resolution | Accuracy(℃) |
| J | -200℃~0℃ 0℃~1200℃ |
0.1℃ |
1.0 0.7 |
| K | -200℃~0℃ 0℃~1370℃ |
0.1℃ |
1.2 0.8 |
| T | -200℃~0℃ 0℃~400℃ |
0.1℃ |
1.2 0.8 |
| E | -200℃~0℃ 0℃~950℃ |
0.1℃ |
0.9 0.7 |
| R | -20℃~0℃ 0℃~500℃ 500℃~1750℃ |
0.1℃ |
2.5 1.8 1.4 |
| S | -20℃~0℃ 0℃~500℃ 500℃~1750℃ |
0.1℃ |
2.5 1.8 1.5 |
| B | 600℃~800℃ 800℃~1000℃ 1000℃~1800℃ |
0.1℃ |
2.2 1.8 1.4 |
| N | -200℃~0℃ 0℃~1300℃ |
0.1℃ |
1.5 0.9 |
| Pt100 | -200℃~850℃ | 0.1℃ | 0.3 |
| Pt1000 | -200℃~650℃ | 0.1℃ | 0.15 |
| Cu50 | -50℃~150℃ | 0.1℃ | 0.5 |
| Cu100 | -50℃~150℃ | 0.1℃ | 0.25 |
1. New types of thermocouples and thermal resistors can be added according to customer needs.
2. The thermal resistance index is a four wire system index.