RF power meter GX2BB01A
The GX2BB01 series RF power meter is a new generation of detector diode power meter based on digital signal processing technology. It has a built-in 5
Product details

GX2BB01A RF power meter


1 overview

GX2BB01seriesRF power meter(Microwave power meter)It is a new generation of detection diode power meter based on digital signal processing technology, with built-in50MHzCalibration source provides power reference, which can calibrate the host and power sensor, with automatic calibration and zeroing, built-in10A storage/retrieval register for users to store and call data;GX2BB01seriesRF power meterMicrowave power meterMatching YS11801XSeries power sensors have the advantages of wide frequency range, large dynamic range of power measurement, fast measurement speed, and high measurement accuracy.

GX2BB01seriesRF power meterMicrowave power meter Can be manually operated through the front panel orUSBThe interface control completes the calibration of the power indicator and the measurement of the power level,GX2BB01seriesRF power meterMicrowave power meterSuperior performance can fully meet the needs of fields such as radar, communication, broadcasting, and televisionRFMeasurement of microwave power.

II Microwave power measurement technology

In the DC and low-frequency fields, power measurement can be achieved by measuring voltage and current, but when the signal frequency is in the tensMHzEven hundredsGHzAt that time, traditional methods could no longer meet the needs of power measurement, so low-frequency measurement methods lost their own significance, and were replaced by electromagnetic field theory. Microwave power meters are a better means of solving microwave power measurement.

The microwave power meter consists of two parts: one is the power sensor (power probe), which directly determines the bandwidth and dynamic range of microwave power measurement. The other part is the power indicator (referred to as the power meter), which is used to set and display the measurement method, menu functions, and power measurement results of the power meter.

Power unit:

Generally, watts are used to represent the unit of power, with symbolsW”It can also derive units such as kilowatts, megawatts, milliwatts, and microwatts, which are also common in the DC and low-frequency fields. There is another commonly used unit for power: decibels and milliwatts(dBm

It can be calculated according to the following formula:

A =10log( P/ P0) (dBm)

WherePP0All are symbols representing power, with power units in milliwatts(mW)Among themP0defined as1mWThe reference

Power.

Example:

1mWThe power, converted by the above equation, is0 dBm

10mWThe power, converted by the above equation, is10 dBm

0.1mWThe power, converted by the above equation, is -10 dBm

The unit of dBm is a widely used power unit in the microwave field.

3、 Technical specifications of microwave power meter

Frequency range

10MHz ~ 18GHz(Depending on the power sensor configuration)

power range

-50dBm ~ +20dBm10nW~100mW)(Depending on the power sensor configuration)

Zero setting

±10nW

display resolution

The logarithmic method is1-0.01dB

50MHz/1mWPower calibration source

Power accuracy ±2%

temperature range

0 ~ 40

Storage temperature

-40 ~ 70

External dimensions

330×260×105mm

weight

make an appointment3.5kg

4、 Main technical indicators of power sensors

model

YS11801A

YS11801B

YS11801C

Frequency range

10MHz12.4 GHz

10MHz18 GHz

10MHz18 GHz

power range

-40dBm+20dBm

100nW~100mW

-40dBm+20dBm

100nW~100mW

-50dBm+20dBm

10nW~100mW

*High power

+23 dBm

+23 dBm

+23 dBm

Power linearity

-20dBm+20dBm

±0.21dBm/10dB

Other ±0.09dB/10dB

-20dBm+20dBm

±0.21dBm/10dB

Other ±0.09dB/10dB

-20dBm+20dBm

±0.21dBm/10dB

Other ±0.09dB/10dB

VSWR

Full frequency band less than or equal to1.29

Full frequency band less than or equal to1.29

Full frequency band less than or equal to1.29

Input connector form

N(M)

N(M)

N(M)

temperature range

0 ~ 40

0 ~ 40

0 ~ 40

Storage temperature

-40 ~ 70

-40 ~ 70

-40 ~ 70

Dimensions (excluding wire length)

weight

make an appointment0.4kg

make an appointment0.4kg

make an appointment0.4kg

Note: The power meter is matched with a power sensor and equipped with our company's precision attenuator, which can be extended to a series of medium to high power testers

Its maximum power can reach1W2W10W100W200W300W...... Wait.

5 Control Description and Function Key Introduction

5.1Control description (as shown below)

5.2Function Introduction

Panel Introduction

function

1)Power switch

Used to turn on and off the use of this instrument.

2LEDliquid crystal display

Used to set and display the measurement method, menu status, and power measurement results of the power meter.

3)Keyboard

Set and determine the measurement mode of the power meter, store and call the setting results. (The specific introduction of function keys is shown in the table below).

4)Power reference

pass throughNType connector outputs one1mW0dBm50MHzThe reference signal is used to calibrate the probe and host system. When the LED next to the connector lights up, it indicates that the reference source is turned on.

5)Sensor interface

Power sensor access point

6USBinterface

USBThe interface completes the calibration of the power indicator and the measurement of power level through remote control operation.

7)Socket fuse

Used to protect instruments

8)Power plug

AC 95220V45 55Hz

keyboard

function

ZERO

This keyboard has zero adjustment function. Press this key to enter the zero adjustment menu

CAL

This keyboard is for calibration function. Press this key to enter the calibration menu

MENU

This keyboard is a menu key, press this key to enter its upper sub menu. And menu selection can be made through the 【 ▲▼ 】 keys to be introduced later.

The submenu includes:

AVERAGE】【OFFSET】【FREQUENCY】【dBm/dBr】【SAVE】【CALL】【POWER REF

dBm/Watt

Used to switch display between logarithmic and watt modes

◄►】▲▼

The directional keys are used to select and change parameters;Two function keys

ENTER

Used to confirm the setting results

6、 Usage Introduction

61 precautions

(6.1.1)In useGX2BB01seriesMicrowave power meter (RF power meter)Before proceeding, it is necessary to read the instrument manual and familiarize oneself with the operating methods;

(6.1.2)The instrument should be used within the allowable range of power supply voltage(AC 95220V45 - 55Hz);

(6.1.3)GX2BB01seriesMicrowave power meter (RF power meter)Generally, it can work normally after a few minutes of startup. If precision testing is required, the preheating time is30 minute;

(6.1.4)Before zeroing, the sensor must be suspended or connected to the built-in instrument50MHz / 1mWOn the reference source and turn off the output of the reference source, press【ZERO】Press the key to start zeroing, and the display will show that zeroing is complete. Then, turn on the standard power output to calibrate the power reference.

(6.1.5)In progressmodernThe measurement frequency of the host should be set according to the frequency of the tested device during accurate measurement.

62 Operating instructions

6.2.1Turn on the instrument

Connect the power sensor cable to the signal input port of the host. Plug in the power cord(95245 VAC)Turn on the power switch to initiate the initialization process of the instrument. After initialization is complete, enter the working state. If the sensor is not connected toSENSORThere will be a warning above“NO SENSOR”.

6.2.2Specific operation instructions for the buttons

The front panel has11 A button, including4A function key【ZERO】The 【CAL】The 【MENU】The 【dBm/Watt】The4 A directional key【◄►▲▼ and1 A confirm key【ENTER】.

6.2.3Zero adjustment(ZERO)Calibration(CAL

The instrument ensures the accuracy of the data during the measurement processmodernIt is necessary to frequently perform zeroing and calibration operations on the instrument.

1)Zero adjustment【ZERO

Zeroing is used to calibrate the zero position of a power meter, so that when there is no signal power input to the power sensor, the reading of the power meter is zero (or only a small noise power). During the zeroing process, there will be a waiting process.

It is recommended to zero the power meter in the following situations:

Ⅰ. Before starting up and powering on * * times for power measurement

Ⅱ. When the temperature change exceeds 5 ℃

Ⅲ. When replacing the power sensor (power probe)

Ⅳ. After working for a long time

Before preparing to measure the small signal power (the sensitivity specified by the power sensor is less than -20dBm)Under power conditions, pressZERO "key, instrument's LCDThe LCD display will show“Zero/MENU/ENTER”If it is confirmed to zero, press“ENTER "starts the zero adjustment operation, which is displayed at this time“Zeroing”Waiting for zero adjustment to be completed; If zero adjustment is not required, press“Exit MENU and return to the main interface. (Note: Zero adjustment requires turning off the signal source)

2)Calibration【CAL

Calibration is the process of calibrating the overall gain of the power meter channel and sensor using a 50MHz 1mW (0dBm) reference signal. The purpose of calibration is to set the correct reference calibration factor for the sensor. If using the built-in reference source of the power meter, it is necessary to connect it first and then perform calibration operation. During the calibration process, relative factors such as bias and frequency response are ignored

Press the CAL button to access the instrument's LCDThe LCD display will show“Calibrate/MENU/ENTER If it is confirmed to zero, pressStart the calibration operation with 'ENTER', and it will display“Calibrating”Waiting for calibration to be completed; If calibration operation is not required, press“Exit MENU and return to the measurement interface.

6.2.4Menu(MENU

MENU】The menu mainly provides7 Sub menu functions, including【AVERAGE】The 【OFFSET】The 【FREQUENCY】The

dBm/dBr】The 【SAVE】The 【CALL】The 【POWER REF】.

When setting the corresponding parameters in the system menu, press【MENU】Keyboard, and select the desired option through 【 ▲▼ 】, the selected option has a highlight effect.


6.2.5dBm/Watt

This keyboard is used to switch the measurement units displayed.