Ø main features
l Advanced embedded microcontroller technology;
l High quality imported electrodes, effective alkalization method, and unique automatic stability judgment function ensure the stability and reliability of measurement
l Advanced surface mount technology and integrated design improve cost-effectiveness and reliability
l Exquisite structure, disc installation, all aluminum frame box, beautiful and sturdy, strong anti-interference ability
l Simultaneous display of multiple parameters, rich content, easy to understand
l Automatic or manual temperature compensation
l Two way isolation current output and two way alarm signal
l Large capacity historical data storage and historical curve query function
Ø Main Technical Performance
Ø Electronic unit
Display: 192 × 64 dot matrix LCD, can choose to display Chinese or English
Measurement range: (0.00~9.36) pNa, 0.01ug/L~23.0g/L, (-1000~+1000) mV
Resolution: 0.01pNa, 0.01 μ g/L, 0.1mV
Basic error: ± 0.02pNa
Stability: ± 0.02pNa/24h
Current output: (4-20) mA (two-way isolated output)
Output accuracy: ± 1% F.S
Output load resistance: 800 Ω
Power supply: AC (85-265) V, frequency (45-65) Hz
Power consumption: ≤ 10W
Insulation strength: power supply to ground 1500V/1 minute without arcing
Alarm output: Two channel alarm output, programmable alarm content
Communication: RS485 (optional)
Automatic data storage: can be set to automatically record historical data, historical events, and calibration records
Ø Whole machine
Measurement range: (1.00~8.00) pNa, 0.23ug/L~2300mg/L
Resolution: 0.01 pNa or 0.01 μ g/L
Basic error: ± 0.05pNa or ± 5 μ g/L, take the larger value (within ± 10 ℃ of the calibration temperature)
Response time (T90): 2min (25 ℃)
Temperature sensor: Pt1000
Temperature measurement range: (0.0~99.9) ℃
Temperature measurement accuracy: ± 0.5 ℃
Temperature resolution: 0.1 ℃
Temperature compensation range: (0.1-60.0) ℃ (manual or automatic)
Calibration methods: sample calibration, standard solution calibration
Dimensions: 450mm x 690mm x 190mm (width x height x depth)
Hole size: 410mm × 645mm
Weight: 20kg
Ø sample requirements
Temperature: (5-50) ℃
Pressure: (50-300) kPa
Flow rate: (6-20) L/h
The allowed solid content of the water sample is ≤ 5 microns (no gelatinous substances or oil are allowed)
Ø environmental condition
Environmental temperature: (5-45) ℃
Environmental humidity: ≤ 90% RH (no condensation)
Storage and transportation temperature: (-20~55) ℃ (excluding electrodes, electrodes should be above 0 ℃)