Zhengzhou Qutong Electrical Equipment Co., Ltd
Home>Products>Nitrogen analyzer/protein analyzer
Nitrogen analyzer/protein analyzer
The use of this product in feed analysis: measurement of crude protein content KDN-B nitrogen analyzer - Kai style
Product details

The use of this product in feed analysis: crude protein content

measure

KDN-B typeNitrogen analyzer - Kai typeAlso known as protein analyzer

(Composed of KDN-B type fixed nitrogen distiller and KDN-04 type well digestion furnace)

KDN-B type fixed nitrogen distiller

Model: KDN-B nitrogen analyzer (distiller)
Measurement range: 0.1-200mg nitrogen (0.05% • 95%)
Reproducibility: 1% relative error (including digestion steps)
Recovery rate: 99.5%~100%
Energy consumption: 1000W
Dimensions: 350 * 350 * 730
Weight (KG):
Function and Features: Adopting distilled water to adapt to differences in conductivity and automatically controlling the water inlet and replenishment device of the steam generator

KDN-04 Well type Digestive Furnace

The use of a well type electric furnace core for heating has improved the digestion heat effect. When digesting the sample, weigh the sample (usually less than 1 gram), add concentrated sulfuric acid and catalyst (adding catalyst can accelerate the digestion process and shorten the digestion time). Harmful gases such as CO2 and SO2 that escape during the digestion process are discharged into the sewer through a negative pressure generated by the nitrogen analyzer through a drainage pipe and a three-way drainage system, effectively suppressing the escape of harmful gases. The installation of toxic gas fume hoods in the laboratory is omitted; The entire digestion process of the sample can be completed in about 45 minutes.
(Note: The entire nitrogen analyzer needs to be equipped with a nitrogen distiller, which can be equipped with other models according to actual conditions)
Technical indicators:
Tested varieties: grain, food, dairy products, beverages, soil fertilizers, feed, and other agricultural and sideline products
Measurement range: 0.1-200 μ m nitrogen, 0.05% -95%
Reproducibility: 1% relative error (including digestion steps)
Temperature control method: thyristor voltage regulation and temperature control
Power: Four hole digestion furnace 1000W
Dimensions: Four hole digestion furnace 630 × 205 × 170mm 3
Weight: Four hole digestion furnace 9kg