Product details

Product Introduction

traditionalVickers hardness testerDifferent, this visual seriesVickers hardness testerCancel the use of measuring eyepieces, project the Vickers indentation image onto the operation screen, and measure the indentation diagonal on the screen. The testing process is faster and more intelligent to meet your requirements for efficient, convenient, and accurate hardness testing.


Characteristics of "Aolongxin"

High precision step controlled electric loading technology

1. The loading speed of the main test force is extremely fast, resulting in improved efficiency

2. Real time video measurement for better, more accurate, and stable results

3. High precision sensor coordinationAolongxin Control System

4.Unique force curve chart of "Aolongxin"

5. Add and unloadFaster and more precise loading speed

6.Adopting unique embedded visual technology

7.Simple measurement operation greatly reduces human error

8.Can achieve remote control, easy to operate

9. Extremely fast loading speed: test force loading time2-10 seconds

10.Intelligent control throughout the entire process. Segmented force application, smooth variable speed, high-precision and high-speed step control, etc


Visual system characteristics

Objectivity: The visual system has objectivity, and the analysis results are not affected by subjectivity

High grayscale resolution: average for the human eye64 grayscale, machine vision can reach up to 256 grayscale

Embedded imaging technology replaces traditional measuring eyepieces with cameras+Display screen

Mode, indentation image, directly displayed on the screen

Visual chip type: Industrial grade dedicated high-definition camera

Contrast, brightness, white balance and other adjustments are all available

The visual analysis mode simplifies the manual operation process, reduces personnel fatigue, and makes measurement data more intuitive and convenient

Optimize each step, greatly improve overall performance compared to traditional hardness testers, and save time50%


The advantages of visual system over eyepiece measurement

Directly measure on the surface that needs to be measured without visual error. Traditional hardness testers use eyepieces for measurement, and the scale line and indentation are not on the same plane. Under different viewing angles, there will be visual differences between the scale position and indentation position, resulting in differences in measurement results from different viewing angles and personnel

High measurement accuracy: There is no problem of mechanical eyepiece backlash. The resolution of the mechanical eyepiece spiral micrometer scale isAbout 10 μ m, and there is mechanical hysteresis, so the measurement method needs to be strictly defined. The pixel spacing of the image sensor in the visual system is 7.5 μ m, with relatively high accuracy and good measurement consistency

Multiple measurement methods can be provided for simultaneous two-dimensional measurement, simplifying the measurement process and improving measurement speed

Capable of rectangular tangent measurement method

No visual fatigue, reducing measurement labor intensity. The Aolong visual system adopts65K true color TFT LCD screen, comfortable to observe without fatigue, and can be measured by multiple people at the same time


Technical Specifications

model

MHVS-5V

MHVS-10V

MHVS-30V

MHVS-50V

Test force (kgf)

0.2, 0.3, 0.5, 1, 2, 2.5, 3, 5

0.3, 0.5, 1, 2, 2.5, 3, 5, 10

0.5, 1, 2, 3, 5, 10, 20, 30

1. 2, 3, 5, 10, 20, 30, 50

loading method

Automatic loading and unloading

Test force switching

automatic

Language Selection

Chinese-English translation

Duration Time

5-60s

Pressure head objective lens switching

automatic

Testeable parameters

HV

objective lens

5 times, 20 times

Maximum measurement length

400μm

minimum resolution

0.1μm

Maximum height of sample

100mm

Distance from the center of the pressure head to the body

95mm

Hardness conversion

HVTheHKTheHRATheHRBTheHRCTheHRDTheHRETheHRFTheHRG.TheHRKTheHR15NTheHR30NTheHR45NTheHR15TTheHR30TTheHR45TTheHBW

Touch screen operation panel

Image, force curve, optionsdataLibrary, control, loading startLight adjustmentObjective lens switching

numbervaluedisplay

D1D2 valueTheHVHK instructionshardness valueconversion valueLoad retention timeTest count

light source

5V/3WWhite LED

communication interface

RS 232

execution standard

JJG151,GB/T 4340ISO/DIN6507ASTM E-384

Photographic device

Built in high-definition camera

power supply

AC90-240,50-60Hz

Overall dimensions(Length, Width, Height)

520*260*550mm

Net weight of the machine

About 40kg

Sample Requirements

Although Vickers hardness can measure both softer and harder materials, it also has its own requirements for the sample. Only by selecting appropriate samples can we avoid the errors caused by this and obtain accurate Vickers hardness values.

Appearance requirements for the sample

The surface of the Vickers hardness test sample should be smooth and flat, without oxide scale or impurities, and without oil stains. Generally, the surface roughness parameter of Vickers hardness test specimensRa is not greater than 0.40 μ m, small load Vickers hardness specimens are not greater than 0.20 μ m, and micro Vickers hardness specimens are not greater than 0.10 μ m. (μ m is the unit of surface roughness parameter Ra.). )

Requirements for sample preparation

During the preparation process of Vickers hardness specimens, the influence of factors such as overheating or cold work hardening on surface hardness should be minimized as much as possible.

In addition, for samples with small cross-sections or irregular shapes, such as spherical or conical shapes, it is necessary to embed the samples or use specialized platforms.


Host standard configuration

name

quantity

name

quantity

Cross test bench

1only

Thin film clamping table

1only

Flat mouth clamping table

1only

Fine wire clamping table

1only

screwdriver

2turn

Horizontal adjustment screw

4only

Microhardness block

2block

level gauge

1only

spare fuses

2only

Product User Manual

1copy

Product Qualification Certificate

1copy