VIP member
MC Atomic Absorption Cooling Water System
Product features: MC series cooling water circulation device can generate constant temperature, constant current, and constant pressure cooling water,
Product details
MeichengMCTechnical parameters of series chilled water circulation system
| model | MC-1000C/BP | MC-1600C/BP | MC-2500C/BP | MC-3500C/BP | MC-7000C/BP | MC-8000 | MC-12000 |
| Refrigeration capacity | 1000W | 1600W | 2500W | 3500W | 7000W | 8000W | 12000W |
| Temperature range |
5~35℃
|
5~35℃
|
5~35℃
|
5~35℃
|
5~35℃
|
5~35℃
|
5~35℃
|
| Temp Accuracy | ± 0.1℃ | ± 0.1℃ | ± 0.1℃ | ± 0.1℃ | ± 0.1℃ | ± 1℃ | ± 1℃ |
| return difference | 0 | 0 | 0 | 0 | 0 |
0℃-4℃
|
0℃-4℃
|
| control mode | PID | PID | PID | PID | PID | digital control | digital control |
| compressor | Variable frequency type | Variable frequency type | Variable frequency type | Variable frequency type | Variable frequency type | Start stop type | Start stop type |
| Water tank capacity | 15L | 15L | 15L | 20L | 20L | 80L | 100L |
| Installation method | Integrated mobile type | Integrated mobile type | Integrated mobile type | Split fixed type | Split fixed type | Split fixed type | Split fixed type |
| Input power | 850W | 850W | 850W | 1300W | 2300W | 3300W | 4800W |
| Power requirements | 220V 10A | 220V 10A | 220V 10A | 220V 10A | 220V 10A | 380V | 380V |
| pumping flow |
6L/min
|
6L/min
|
6L/min
|
6L/min
|
6L/min
|
6L/min
|
6L/min
|
| Equipment size W×D×H (mm) | 360×350×540 | 700×485×780 | 700×485×780 |
Indoor unit710×445×690
Outdoor800×280×600
| Indoor unit 710×445×690
Outdoor1020×340×830
| Indoor unit 420×650×800
Outdoor900×310×840
| Indoor unit 420×700×1040
Outdoor900×310×840
|
| Outlet pressure | 7bar | 7 bar | 7bar | 7bar | 7 bar | 7 bar | 7bar |
| notes |
1MeichengMCThe series of chilled water circulators is a national "15th Five Year Plan" scientific and technological research project, and currently holds a position in the field of variable frequency refrigeration in China.7000WThe following is a variable frequency controlled chilled water circulation system.
2The listed configurations are standard, and a booster pump and water tank can be selected according to the actual needs of the user.
| ||||||
Structural characteristics
Exquisite structure and beautiful appearance
Durable and sturdy, capable of long-term continuous operation
Equipped with a digital thermostat to automatically control the water temperature
Easy to connect with the equipped instrument and no water leakage
The all-in-one mobile device is equipped with casters, making it easy to move
Choose different high-quality imported silent pumps with reasonable structure, low noise, good performance, and high reliability
The installation and use are very convenient, without the need for special installation or pipeline connection. Simply connect to the analytical instrument to be cooled, then turn on the power, set the temperature, and press the "run" switch
Technical performance characteristics
■ Variable frequency characteristics
1) cooling water temperature5-35Within the temperature range of ℃, it can be adjusted according to requirements. By changing the frequency of the compressor, the input and output power of the refrigeration system can be automatically adjusted to achieve temperature control, energy saving, and energy-saving compared to the start stop type with the same refrigeration capacity30%the above
2) The key components of the refrigeration system, such as compressors and circulating fans, are imported products. When the temperature is balanced, the compressor runs at low speed with minimal noise; Adopting imported circulation pump, with large circulation flow and maximum pressure reaching7.0barabove; Adjustable pressure
3) Micro processing chip control, high control accuracy, stable and reliable performance, andPIDCombining temperature control technology to achieve precise temperature control(±0.1℃)
4)Input power during compressor start-up20Within seconds, gradually increasing from small to large“soft start”This avoids the high current surge during the start-up of a constant speed compressor and extends the lifespan of the entire machine
5)By changing the frequency of the compressor to achieve temperature control, the input and output power of the refrigeration system can be automatically adjusted
■ Start stop characteristics
1) Real time display of temperature, pressure, and water level
2)Using soft water medium significantly improves working life and reliability
3) Dynamic constant temperature control system, fast response, small constant temperature fluctuations
4) Key components of the refrigeration system, such as compressors and circulating fans, are imported products
5) Adopting imported circulation pump, with large circulation flow and adjustable pressure; Optional circulation pump with higher outlet pressure
4) Key components of the refrigeration system, such as compressors and circulating fans, are imported products
5) Adopting imported circulation pump, with large circulation flow and adjustable pressure; Optional circulation pump with higher outlet pressure
■ Split type characteristics
Split type chilled water cycle unit is designed for high-power instruments, where the refrigeration system and heating system are installed separately indoors and outdoors, reducing noise and the impact on indoor temperature. Thereby reducing the footprint of the laboratory
Instrument Introduction
MCSeries Cooling Water Recirculators1995Since its launch in the market, the product has undergone multiple improvements and is now very mature and reliable, highly welcomed by users in the market.MCThe series of chilled water circulators can generate constant temperature, constant current, and constant pressure cooling water, suitable for various instruments and equipment that require water cooling and constant temperature. MeichengMCThe series of chilled water circulators is a national "15th Five Year Plan" scientific and technological research project, and currently variable frequency refrigeration is in a prominent position in China.
Main application areas
◆Analytical instruments: graphite furnace atomic absorptionICP、ICP-MSElectron microscope, mass spectrometerX-Diffraction instrumentX-Fluorescence spectrometer, infrared spectrometer, viscometer, polarimeter, fermentation device, rotary evaporator
◆Vacuum equipment and medical: molecular pumps, diffusion pumps, cooling blanketsXOptics, nuclear magnetic resonance, accelerators
◆Industrial equipment and others: laser equipment, vacuum coating equipment, biopharmaceuticals, molds, machine tools, vacuum furnaces, spot welding machines, reaction vessels
Online inquiry