Instrument Introduction
LD3001 is a catalytic methane/non methane total hydrocarbon portable or wall mounted multi-purpose analyzer. It adopts full heating, no cold point high-temperature cycle technology, is equipped with a high-temperature and efficient catalytic device, and uses dual FID detectors to simultaneously detect the concentration values of total hydrocarbons and methane in the sample gas. It can display and calculate the concentration values of total hydrocarbons, methane, and non methane in real time.
Complies with the national standards of "Technical Requirements and Testing Methods for Continuous Monitoring System of Non Methane Total Hydrocarbons in Fixed Pollution Source Waste Gas" (HJ 1013-2018) and "Technical Requirements and Testing Methods for Portable Monitoring Instruments of Total Hydrocarbons, Methane and Non Methane in Environmental Air and Waste Gas" (HJ 1012-2018).
Catalytic FD principle:
The biggest difference between using catalytic oxidation FD method and chromatography-FD method is that instead of using a chromatographic column to separate methane, a methane cutter (high-temperature catalytic oxidation furnace) is used to separate methane.
Instrument principle
Import the sample into a portable catalytic oxidation hydrogen flame ionization detector analyzer, where all hydrocarbons in one route are detected by FID to obtain the total hydrocarbon content. The hydrocarbons in the sample gas (i.e. air) are catalytically oxidized to carbon dioxide and water to obtain the methane content. The difference between the two is the non methane total hydrocarbon content.
Instrument features:
The instrument is highly integrated and scalable, and can be configured with a total hydrocarbon analyzer and a methane/non methane total hydrocarbon analyzer according to demand;
The entire process of high-temperature heating, including system analysis of gas pipelines and gas control components, is carried out in an environment with a high temperature of 190 degrees Celsius, and there are no cold spots during the analysis process;
Under the action of the new catalyst, non methane hydrocarbons do not produce any reaction components, and the catalytic capacity reaches 95%;
Adopting a dual FID detector flow path design, simultaneously detecting total hydrocarbon and methane concentration values in parallel to ensure accuracy and real-time analysis;
Dual channel EPC pressure controller ensures consistency of total hydrocarbon and methane injection quantities;
The instrument is easy to operate, and after preheating, it can continuously detect methane, total hydrocarbons, and non methane concentrations without interruption, achieving real-time detection;
The auxiliary gas source is single, only requiring micro compressed hydrogen gas;
Low maintenance costs, no consumable parts in the instrument, and a catalyst lifespan of over 2 years.
technical indicators
Measure the composition of hydrocarbons
Measurement range 0.1-1000mgC/m3 (CH4)
Detection limit ≤ 0.1mgC/m3
Conversion rate ≥ 95%
Relative standard deviation ≤ 2% (CH4)
Repeatability 1%
Measurement accuracy ≤± 0.2%
Linear error ± 2% (CH4)
Reaction time T90 ≤ 1.5s
Technical Specifications
Measurement method: Catalytic+flame ionization detector (FID)
The FID detector automatically determines the ignition flame status of the detector and has automatic ignition and flame judgment alarm functions
The sampling system is heated throughout the entire process
Sampling flow rate 600mL/min
Hydrogen metal hydride technology
Standard gas methane/propane+air bottom gas
AC power supply AC-220V, instantaneous maximum total power 300w (excluding heating of sampling tube)
technical advantage
Using catalytic FD principle, built-in methane cutter, high temperature throughout the process, accurate data
Unique dual detector configuration, with a measurement response time of less than 3 seconds for non methane total hydrocarbons;
The auxiliary gas source is single, requiring only micro compressed hydrogen gas or solid-state hydrogen gas storage modules;
Dual channel EPC pressure controller ensures consistency of total hydrocarbon and methane injection quantities;
The ∪ disk directly stores all measurement data and instrument status information in E ⅹ cel/cs ∨ format;
Low maintenance costs, no consumable parts in the instrument, and a catalyst service life of over 2 years;
The total weight is 20 kilograms, with a handle design, convenient for single person carrying, and can be measured horizontally or vertically;
Application field
Engine exhaust analysis;
Emission measurement of catalytic, biological, and activated carbon treatment devices
Monitoring of Hydrocarbons in Crude Gas to Clean Gas Conversion
Emissions purification and monitoring of hydrocarbons in chemical clean factories
Workplace permissible concentration measurement
Process monitoring in manufacturing enterprises
Emission monitoring of power plants, incineration plants, and mechanical waste treatment plants
Government environmental protection agencies, environmental monitoring stations, and research institutions