GX2C1B1000 high-power microwave power meter
GX2C1B1000 microwave power meter is a new generation microwave power measurement instrument. GX2C1B1000 Microwave Power Tester Dynamic Range50dB,Detect using a detection diode method. An upgraded product that replaces the original bismuth antimony thin film beachhead power meter with a sensor To complement the combination of modern instruments. GX2C1B1000 Microwave Power TesterUsers can choose withGPIBInterface,RS232perhapsRS485Interface is used to implement transportation control or to form a measurement system together with other instruments.
Technical parameters of GX2C1B1000 high-power microwave power meter:
(1)Frequency range: 20MHz~12.4GHz
(2)Power measurement range:
0.1μw~1000W (alternatively available: 0.1)μw~300W; zero point oneμw~500W; 01μw~800W and other specifications)
(3)Measurement accuracy: inherent error ±5%
(4)Voltage standing wave ratio:S≤1.4
(5)Power benchmark:50MHz(1.00mW±1.5%)
(9)Recorder output:0~1VVoltage, output impedance1KΩ
(10)Display mode:WMethod,dBmMethod anddBMethod.
(11)Interface capability:GPIB、RS232andRS485. (All purchased by users)
(12)Power supply: AC~220V、50Hz、25W
(13)Host dimensions:99mm(tall)×220mm(Width) ×350mm(Deep)
(14)Attachment dimensions: 410mm(tall)×135mm(Width) × 650mm(Deep)
Detailed introduction of GX2C1B1000 high-power microwave power meter
1. GX2C1B1000 high-power microwave power meteroverview
GX2C1B1000 microwave power meter is a new generation microwave power measuring instrument. GX2C1B1000 microwave power meter is an upgraded product of the original GX2C microwave power meter. The dynamic range of GX2C1B1000 microwave power meter is 50dB The GX2C1B1000 microwave power meter uses a detection diode for detection. The GX2C1B1000 microwave power meter can be equipped with GPIB, RS232, or RS485 interfaces for distance control or measurement system composition. GX2C1B1000 microwave power meter is a microwave power measuring instrument. The GX2C1B1000 microwave power meter consists of a GX2C1B1000 power indicator and a 11801 power sensor. And users can choose RS232 interface or GPIB interface. The GX2C1B1000 microwave power meter can be used for microwave power measurement in factories and laboratories.
2. Technical parameters of GX2C1B1000 high-power microwave power meter
2.1 Frequency range: 20MHz to 12.4GHz
2.2 Power measurement range:
0.1μw~1000W (alternatively available: 0.1)μw~300W; zero point oneμw~500W; zero point oneμw~800W and other specifications)
2.3 Measurement accuracy: inherent error ± 5%
2.4 Voltage standing wave ratio: S≤1.4
2.5 Can withstand an average power of: 1 minute of excess power test.
2.6 Power reference: 50MHz
1.00mW±1.5%
2.7 Noise drift: ± (0.1uW+1 word) within any minute interval
2.8 Recorder output: 0-1V voltage, output impedance 1K Ω.
2.9 Display mode: W mode, dBm mode, and dB mode.
2.10 Power Supply: AC mains~220V, 50Hz, 20W
2.11 Overall dimensions of the host: 99mm (height) × 220mm (width) × 350mm (depth)
2.12 Attachment dimensions: 410mm (height) x 135mm (width) x 650mm (depth)
2.13 Weight: Approximately 4Kg (host); Approximately 36Kg (attached).
The role of the adjustment and control mechanism of GX2C1B1000 high-power microwave power meter
Front and rear panel adjustment control mechanism
3.1 LED display window
Display power measurement readings.
3.2 power switch
When the power meter is connected to the power supply, if the indicator light is red, it indicates that the instrument is in standby mode. By switching the power switch and turning the indicator light green, the instrument is in working condition.
3.3 keyboard
The keyboard includes 8 keys, including mode, zero adjustment, calibration, setting, storage, retrieval, system, and Enter. For specific operation methods, please refer to the panel operation terms.
3.4 heading
As a linear indicator of RF signals, it reflects the trend of digital display.
3.5 Sensor input port
The sensor inputs a signal through this input port.
3.6 Reference power output port
This output port is turned on (indicator light on) and off by the reference key (OSC). Output 50MHz when turned on
A reference power signal of 1.00mW.
3.7 Recorder output port
Output voltage of 0-1V, output impedance of 1K Ω, BNC connection.
3.8 Power sockets
3.9 Fuse holder
4. Working principle of GX2C1B1000 high-power microwave power meter
4.1 Whole machine block diagram
4.2 Principles of each unit circuit
a. Power sensor
The power sensor is composed of a chopper and a linear correction circuit. A chopper is a series chopper composed of highly sensitive and low-noise MOS-FET switching transistors. Its function is to modulate the DC signal input through the power probe into an equivalent AC signal, which is more convenient for the processing of the subsequent amplification circuit. The linear correction circuit is a spreadsheet that stores sensor parameters and is processed by the CPU.
b. Preamplifier
A preamplifier is a hybrid amplifier composed of high gain transistors and high impedance, low barrier operational amplifiers. The AC gain of this amplifier is about 600 times, which utilizes the low noise of transistors and the high gain characteristics of operational amplifiers.
c. Attenuation network and main amplifier
The attenuation network is composed of analog switches and resistors, and is controlled by a central processing unit. It determines the range conversion. The main amplifier is composed of programmable gain amplifiers. It can be adjusted to an appropriate gain under the control of the central processing unit.
d. Amplify network and analog-to-digital converter
The amplification network is composed of high impedance and low barrier operational amplifiers X1, X10, and X100, respectively, and generates signals of three different ranges. Three signals enter the multi-channel, high-speed analog-to-digital converter simultaneously, and the central processor reads the corresponding data according to different ranges.
e. Zeroing network
Zeroing network is composed of digital to analog converters and corresponding devices. When the instrument is performing a zero adjustment operation, the data read from the analog-to-digital converter by the central processing unit is analyzed accordingly and the digital to analog converter provides the opposite signal to cancel out the zero point error in the circuit and achieve the zero adjustment effect.
f. Chopper driver
Amplify the current of the 220Hz square wave signal supplied to the chopper.
g. Recorder output circuit and meter drive circuit
The output circuit of the recorder and the driving circuit of the electric meter are composed of digital to analog converters and corresponding operational amplifiers. The central processing unit drives this part of the circuit based on the results of each measurement.
h. Central Processing Unit
The central processing unit adopts a high-performance 8-bit microprocessor. It manages the operation of the entire instrument.
i. Panel Circuit
The panel circuit consists of an LED display window and a user keyboard. It is the main tool for human-machine dialogue during manual operation.
j. 50MHz reference source
Provide a 50MHz, 1mW power reference signal. When the sensor is connected to a 50MHz reference source, the instrument can be calibrated by 1mW through the "calibration" operation.
5. Structural characteristics of GX2C1B1000 microwave power meter
The external dimensions of the indicator part of GX2C1B1000 microwave power meter are 220mm × 99mm × 350mm (b × h × d), and the display part of GX2C1B1000 microwave power meter is an LED display screen. The front panel buttons are made of conductive rubber technology, and their control methods are all operated through button menu. Connect and form a structure that is easy to maintain and produce. The channel plate adopts an electrical shielding structure to reduce the interference of stray electromagnetic fields. The rear panel of GX2C1B1000 microwave power meter consists of a power input socket and a recorder output socket.
6. Instructions for using GX2C1B1000 high-power microwave power meter (see product manual)
Related products:YS-2499A(B)seriesDual channel pulse high-power meter;GX2BB01Series RF power meter; YS1125 (B) RF signal generator; YS3892 frequency selective amplifier; PXSeries microwave frequency meter;FYS3000Complete waveguide testing systemAnd accessories;RCQ1AMicrowave leakage energy tester;SHK-4(3)Variable RF attenuator;BSeries fixed RF attenuator;TJ8-4, TJ8-3, TJ8-5, TJ8-2 series coaxial crystal RF detectors