YS-2499A Dual Channel Waveform Pulse Power Meter
1. overview
YS-2499A dual channel waveform pulse power meter (waveform display microwave power meter) is a newly upgraded multifunctional dual channel digital peak power meter. YS-2499A dual channel waveform pulse power meter (waveform display microwave power meter) adopts4.7The inch color LCD screen serves as the display interface, and high-quality conductive rubber serves as the human-machine interaction interface. The YS-2499A dual channel waveform pulse power meter (waveform display microwave power meter) has the following characteristics:
(1)Capable of measuring peak power, pulse power, continuous wave power, etc;
(2)Can display the waveform of a single pulse for two channels separately;
(3)Can be done throughMarkerMeasure the peak power value at any position within the pulse;
(4)Rich communication methods, includingLANMouth(Standard configuration for thisFunction not activated)USB (optional)、RS232Serial port, etc;
(5)Equipped with manual zeroing and power value correction functions;
The YS-2499A dual channel waveform pulse power meter (waveform display microwave power meter) can be operated by an external deviceTwo channels measure the incident power separatelyAnd reflected powerThus, corresponding standing wave ratio and reflection system can be obtained, and continuous wave microwave power can also be measured.YS-2499A dual channel waveform pulse power meter (waveform display microwave power meter)Used for standing wave ratio monitoring of various microwave high-power systems. YS-2499A dual channel waveform pulse power meter (waveform display microwave power meter)correctThe upgraded product of YS (M) 2498 dual channel pulse RF power meter.Users can choose specific directional couplers to expand the power measurement rangeIt can be measured up to 100MW
2. Technical Parameter
2.1 Frequency range: 50MHz – 4GHz(Frequency range can be customized according to requirements, up to 12.4GHz)
2.2 Power measurement range:-30dBm~20dBm(1μw ~ 100mW)(Customizable power range according to requirements)
2.3 Display:LCDIndicating power, which can beWattsperhapsdBm
2.4 Power indication accuracy:≤±0.5dB
2.5 Communication interface:LAN(Standard configuration for thisFunction not activated)USB (optional)、RS232
2.6 User selected interface: alarm interface(BNC)
2.7 Power detection output port:0-2V
2.8 Input connector:Ntype
2.9 Power Supply:AC110V-240V 50Hzperhaps60Hz
2.10 Host size:220mm×120mm×280mm(Length x Width x Depth)
pour:Ys2499AAs a universal frequency band measuring instrument, the measurement frequency band can be customized according to the actual needs of users,Expand the power range,And other customized functional services can be provided according to actual situations.
3. Adjust the control mechanism
3.1 LCDdisplay
Display waveform, power, and related parameters
3.2 Power sockets and switches
When the power is plugged in and the switch is turned on, the instrument is powered on
3.3~3.4 Keyboard
Instrument workers include11A button, which are4A function key4Direction keys, zero adjustment keys, calibration keys, and menu keys
3.5 AChannel input port
Connect through this input portAChannel input power (generally considered as incident power)
3.6 BChannel input port
Connect through this input portBChannel input power (generally considered as reflected power)
3.7 LANopening
3.8 RS232interface
3.9 External trigger signal input interface
The synchronous trigger signal for this input and power input interface
3.10 Reflective alarm interface
default outputTTLHigh level, meets the alarm condition outputTTLLow Level
4. Introduction to Screen Display Parameters
4.1 △T
cursor1With cursor2Horizontal spacing (i.e. the time difference between two pointers)
4.2 Mpos
Location of the main time base
4.3 △CHA
AChannel cursor1And cursor2The difference in peak power corresponding to the location
4.4 △CHB
BChannel cursor1And cursor2The difference in peak power corresponding to the location
4.5 Avg(A)、Avg(B)
A、BChannel pulse power
4.6 Peak
A、BPeak power of the channel
4.7 Menu Area
5.Usage
4.1 Precautions
a. Before using this instrument, one should first read the instrument manual and become familiar with the operating methods;
b. This instrument usually works normally after a few minutes of startup. For precision testing, the preheating time is 15 minutes.
4.2 Button Introduction
There are a total of 11 buttons on the front panel of YS-2499A, including 4 function keys, menu key, zero adjustment key, calibration key, and four directional keys. There is a power button at the power input interface on the back panel that controls the power on/off.
a. Function keys
YS-2499A has 4 function keys used to select submenus under each menu page.
b. Menu key
This key is mainly used for switching between menu pages.
c. Zero adjustment key
This key is mainly used for zeroing instruments
d. Correction key
This key is used for instrument calibration
e. Direction keys
YS-2499A has four directions: up, down, left, and right. The up and down keys are mainly used for numerical addition and subtraction, while the left and right keys are mainly used for cursor movement under submenus.
4.3 Menu Introduction
YS-2499A has set up 5 menus, namely Menu1 Menu2、Menu3、Menu4、Meau5, They switch through menu keys, pressing the corresponding function key on the right side of each submenu, and then using the up, down, left, and right arrow keys to operate the submenu or set the corresponding parameters.
a、Menu1: The four submenus under this menu are mainly used to set parameters related to power measurement
(1)、FREQUENCYSet frequency compensation and compensation range0.05GHz ~ 18GHz
(2)、OFFESETSet corresponding channel power compensation, compensation range-99dB ~ +99dB
(3)、DISPLAYSwitching the power display mode can be done atWattsanddBmSwitching between
b、Menu2The four submenus under this menu are mainly used to set parameters related to waveform display
(1)、REF LEVEL: Set the waveform display area * large amplitude value, that is, the power value represented by the top power line *
(2)、TRIG LEVELSet the display waveform to trigger the power level
(3)、TRIG MODESet the waveform triggering method for display, includingAuto、Normal、Single
(4)、SOURCE: Select trigger source, includingCHA、CHB(A、BSignal as trigger)EXIT
(External input trigger signal)
c、Menu3The four submenus under this menu are mainly used to set parameters related to waveform display
(1)、SCAN TIME: Set the time represented by each horizontal grid in the waveform display area
(2)、XPOSTIONMain Time Base Position
(3)、Marker1: Set cursor1where
(3)、Marker2: Set cursor2where
d、Menu4The sub menu of this menu is mainly used for settingA、BChannel alarm power level
e、Menu5The sub dishes in the menu are mainly used for settingLANMouth andRS232Interface related parameters
6.Communication Interface Protocol
USBInterface: (This)USBFor VirtualUSBInterface, similar toRS232Communication format)
|
|
function |
instruction |
*range |
|
1 |
Frequency setting |
ASA_[FREQ](**.**GHz) |
09.30~09.50 |
|
2 |
offset setting |
ASA_[OFFSET](±**.**,±**.** dB) |
-99.99~+99.99 |
|
3 |
Trigger level |
ASA_[TRIG](±**.** dBm) |
-30.00~+20.00 |
|
4 |
Trigger Mode |
ASA_[MODE](*) |
0,1,2(auto,normal,single) |
|
5 |
trigger channel |
ASA_[CH_X](*) |
0,1,2(CHA,CHB,EXT) |
|
6 |
Scan time |
ASA_[TIME](**) |
2~17 |
Network communication protocol (following)ModBusProtocol format)
Set instruction format
|
|
function |
instruction |
“_”meaning |
|
1 |
Frequency setting |
41 53 41 10 A1 02 _ _CRCH CRCL |
2One byte,50MHz~4G Range: 0x00 0x05~0x01 0x09 |
|
2 |
offset setting |
41 53 41 5F A2 04 _ _ _ _ CRCH CRCL |
front2One isCHA after2One isCHB defined as2Bytunsigned number |
|
3 |
Trigger level |
41 53 41 5F A3 02 _ _ CRCH CRCL |
Defined as unsigned2individualBytunsigned number |
|
4 |
Trigger Mode |
41 53 41 5F A4 01 _ CRCH CRCL |
0,1,2(auto,normal,single) |
|
5 |
trigger channel |
41 53 41 5F A5 01 _ CRCH CRCL |
0,1,2(CHA,CHB,EXT) |
|
6 |
Scan time |
41 53 41 5F A6 01 _ CRCH CRCL |
2~17 |
Return instruction format:
The system automatically returns measurement parameters:CHA,CHBofPeakprice
41 53 41 5F B1 04 _ _ _ _ CRCH CRCL
explain:
Data header: 41 53 41 5F
Function code:B1
Data length:04
data
16Bit verification code:CRCH CRCL
The format for setting and returning instructions is the same.
Related products:GX2BB01 series RF power meter; YS1125 (B) RF signal generator; YS3892 frequency selective amplifier; PXSeries microwave frequency meter;FYS3000Complete waveguide testing systemAnd accessories;RCQ1AMicrowave leakage energy tester;SHK-4(3)Variable RF attenuator;BSeries fixed RF attenuator;TJ8-4, TJ8-3, TJ8-5, TJ8-2 series coaxial crystal RF detectors