Dual channel pulse power meter YS-2499AX
YS-2499AX dual channel pulse power meter is a newly upgraded multifunctional dual channel digital peak power meter, with high-quality conductive rubbe
Product details

YS-2499AX Dual Channel Pulse Power Meter

1. overview

YS-2499AX Dual Channel Pulse Power Meter (Dual Channel Waveform Microwave Power Meter) is a newly upgraded multifunctional dual channel digital peak power meter. The YS-2499AX dual channel pulse power meter (dual channel waveform microwave power meter) uses a 4.7-inch color LCD display screen as the display interface, and high-quality conductive rubber as the human-machine interaction interface. The YS-2499AX dual channel pulse power meter (dual channel waveform microwave power meter) has the following characteristics:

(1) Capable of measuring peak power, pulse power, continuous wave power, etc;

(2)Can display the waveform of a single pulse for two channels separately;

(3) The peak power value at any position within the pulse can be measured using a marker;

(4) Rich communication methods, including LAN port (this feature is not enabled as standard), USB, RS232 serial port, etc;

(5)Equipped with manual zeroing and power value correction functions

The YS-2499AX dual channel pulse power meter (dual channel waveform microwave power meter) can be operated through an external deviceTwo channels measure the incident power separatelyAnd reflected powerThus, corresponding standing wave ratio and reflection system can be obtained, and continuous wave microwave power can also be measuredThe YS-2499AX dual channel pulse power meter (dual channel waveform microwave power meter) is also used for standing wave ratio monitoring of various microwave high-power systems. The YS-2499AX dual channel pulse power meter (dual channel waveform microwave power meter) is an upgraded product of the YS (M) 2498 dual channel pulse RF power meter.Users can choose a specific directional coupler to expand the power measurement range, which can measure up to 100MW

2. Technical Parameter

2.1 Frequency range: 50MHz – 18GHz(Customize frequency range according to specific requirements)

2.2 Power measurement range: 1 μ w~800mW(Customize power range according to requirements)

2.3 Display: LCD indicates power, which can be in Watts or dBm

2.4 Power indication accuracy: ≤ ±0.5dB

2.5 Communication interface: LAN (this function is not enabled as standard), USB (optional) RS232

2.6 User interface selection: Alarm interface (BNC)

2.7 Power detection output port:0-2V

2.8 Input connector: N-type

2.9 Power Supply: AC110V-240V 50Hz or 60Hz

2.10 Host size: 220mm x 120mm x 280mm (length x width x depth)

Note:Measurement frequency bands can be customized according to the actual needs of users,Expand the power range,And other customized functional services can be provided according to actual situations.

3. Adjust the control mechanism

3.1 LCD Display Screen

Display waveform, power, and related parameters

3.2 Power sockets and switches

When the power is plugged in and the switch is turned on, the instrument is powered on

3.3~3.4 Keyboard

The instrument operator includes 11 buttons, including 4 function keys, 4 directional keys, zero adjustment key, calibration key, and menu key

3.5 A channel input port

Connect the input port to channel A for input power (generally considered as incident power)

3.6 B-channel input port

Connect the B channel input power (generally considered as reflected power) through this input port

3.7 LAN port

3.8 RS232 interface

3.9 External trigger signal input interface

The synchronous trigger signal for this input and power input interface

3.10 Reflection alarm interface

Default output TTL high level, meet alarm conditions output TTL low level

4. Introduction to Screen Display Parameters

4.1 △T

Horizontal distance between cursor 1 and cursor 2 (i.e. time difference between the two cursor)

4.2 Mpos

Location of the main time base

4.3 △CHA

The difference in peak power corresponding to the positions of cursor 1 and cursor 2 in channel A

4.4 △CHB

The difference in peak power corresponding to the positions of cursor 1 and cursor 2 in channel B

4.5 Avg (A), Avg (B)

A. B-channel pulse power

4.6 Peak

A. Peak power of channel B

4.7 Menu Area


5.Usage

4.1 Precautions

a. Before using this instrument, one should first read the instrument manual and become familiar with the operating methods;

b. This instrument usually works normally after a few minutes of startup. For precision testing, the preheating time is 15 minutes.

4.2 Button Introduction

YS-2499AXThere are a total of 11 buttons on the front panel, including 4 function keys, menu key, zero adjustment key, calibration key, and four directional keys. There is a power button at the power input interface on the back panel that controls the power on/off.

a. Function keys

YS-2499A has 4 function keys used to select submenus under each menu page.

b. Menu key

This key is mainly used for switching between menu pages.

c. Zero adjustment key

This key is mainly used for zeroing instruments

d. Correction key

This key is used for instrument calibration

e. Direction keys

YS-2499AX has four directions: up, down, left, and right. The up and down keys are mainly used for numerical addition and subtraction, while the left and right keys are mainly used for cursor movement under submenus.

4.3 Menu Introduction

YS-2499AX has set up 5 menus, namely Menu1 Menu2、Menu3、Menu4、Meau5, They switch through menu keys, pressing the corresponding function key on the right side of each submenu, and then using the up, down, left, and right arrow keys to operate the submenu or set the corresponding parameters.

a、Menu1: The four submenus under this menu are mainly used to set parameters related to power measurement

(1)、FREQUENCY: Set frequency compensation, compensation range 0.05GHz~18GHz

(2)、OFFESET: Set corresponding channel power compensation, compensation range -99dB~+99dB

(3)、DISPLAY: Switching power display mode, able to switch between Watts and dBm

b、Menu2: The four submenus under this menu are mainly used to set parameters related to waveform display

(1)、REF LEVEL: Set waveform display area * large amplitude value, i.e. the power value represented by the top power line *

(2)、TRIG LEVEL: Set display waveform trigger power level

(3)、TRIG MODE: Set the waveform triggering method for display, including Auto Normal、Single

(4)、SOURCE: Select trigger sources, including CHA and CHB (A and B signals as triggers) EXIT

(External input trigger signal)

C Menu3: The four submenus under this menu are mainly used to set parameters related to waveform display

(1)、SCAN TIME: Set the time represented by each horizontal grid in the waveform display area

(2)、XPOSTION: Main time base position

(3)、Marker1: Set the position of cursor 1

(3)、Marker2: Set the position of cursor 2

d、Menu4: The sub menu of this menu is mainly used to set the alarm power level of channels A and B

e、Menu5: The sub menu is mainly used to set parameters related to LAN and RS232 interfaces


6.Communication Interface Protocol

USB interface: (This USB is a virtual USB interface, similar to RS232 communication format)

function

instruction

*range

1

Frequency setting

ASA_[FREQ](**.**GHz)

09.30~09.50

2

offset setting

ASA_[OFFSET](±**.**,±**.** dB)

-99.99~+99.99

3

Trigger level

ASA_[TRIG](±**.** dBm)

-30.00~+20.00

4

Trigger Mode

ASA_[MODE](*)

0,1,2auto,normal,single

5

trigger channel

ASA_[CH_X](*)

0,1,2CHA,CHB,EXT

6

Scan time

ASA_[TIME](**)

2~17

Ethernet communication protocol (following ModBus protocol format)

Set instruction format

function

instruction

_meaning

1

Frequency setting

41 53 41 10 A1 02 _ _CRCH CRCL

2One byte,50MHz~4G

Range:

0x00 0x05~0x01 0x09

2

offset setting

41 53 41 5F A2 04 _ _ _ _ CRCH CRCL

front2One isCHA

after2One isCHB

defined as2Bytunsigned number

3

Trigger level

41 53 41 5F A3 02 _ _ CRCH CRCL

Defined as unsigned2individualBytunsigned number

4

Trigger Mode

41 53 41 5F A4 01 _ CRCH CRCL

0,1,2auto,normal,single

5

trigger channel

41 53 41 5F A5 01 _ CRCH CRCL

0,1,2CHA,CHB,EXT

6

Scan time

41 53 41 5F A6 01 _ CRCH CRCL

2~17

Return instruction format:

The system automatically returns the peak values of measurement parameters CHA and CHB

41 53 41 5F B1 04 _ _ _ _ CRCH CRCL

explain:

Data header: 41 53 41 5F

Function code: B1

Data length: 04

data

16 bit verification code: CRCH CRCL

The format for setting and returning instructions is the same.


Related products:GX2BB01 series microwave power meter; YS1125 (B) microwave signal source; YS3892 frequency selective amplifier; PXSeries microwave frequency meterFYS3000Complete waveguide testing systemAnd accessoriesRCQ1AMicrowave leakage energy tester;SHK-4(3)Variable RF attenuator;BSeries fixed RF attenuatorTJ8-4, TJ8-3, TJ8-5, TJ8-2 series coaxial crystal RF detectors