1Purpose:
Soil is composed of three phases: solid phase, liquid phase, and gas phase, and their ratio is called the three-phase ratio. This instrument measures this ratio.
IIPrinciple:
The measurement principle is based on Bohr's lawP × V=K. Under air compression conditions, the volume of air is compressed to half, and the air pressure will double. If the pressure of a compressed gas with a known volume can be measured, the unfilled volume of liquid and gas can be calculated. When the temperature rises, the air pressure will increase, so it should be measured in a place where the temperature is relatively constant.

IIITechnical Specifications:
1. Size:W350× D300 × H180mm
2. Measurement range:0~100ml
3. Measurement resolution:0.01ml
4. Measurement accuracy:±0.5ml
5. Weight:7kg
6. Power Supply:100V AC or 12V DC
7. Screen display:4 lines x 20 characters
8. Scope of operation:5~40℃,0~90%RH
9. Internal dustproof design
10. Serial communication:8-pin serial communication, baud rate 19200bps
11. Automatic calibration: automatically calibrate with a test reference body
12. Memory:1000 measurements, can be automatically or manually numbered
13. Display: Can display saved data
14. Data communication: Data can be transmitted throughTransfer DLM software to PC
15. Data saving format:CSV format
16. File Name: Product Serial Number—serial number—latitude and longitude.csv
4form:
Digital soil three-phase measuring instrument, portable case, software, communication cable andRS232 to USB cable, accessories (100ml ring cutter, ring holder, mesh board, test reference body 30ml and 50ml, O-ring, silicone, etc.)
Optional items (recommended for purchase):
l100ml stainless steel ring knife
l1-meter soil drill
lSoil sampler and connecting rod
The instrument is currently only suitable for altitudeFor areas below 500m, models suitable for altitudes of 1000 meters will be launched in the future.
5Place of Origin
Japan