■ Software features:
*This system complies withGB/T 18251-2000National Technical Specifications and Standards of the People's Republic of China.
*This system is used for microscopic image analysis of pigment or carbon black dispersion in polyolefin pipes, fittings, and mixed materials.
*This system can issue national level metrology certificates to ensure the accuracy and effectiveness of the measurement system.
■ Software functions:
1.Software equipped with high-definition, low-noise digital camera,Capable of capturing high-quality images.
2.The software has powerful data processing functions. Based on national inspection standards, the software can select inspection items according to needs. With the help of the software, it can analyze the captured particle images and provide dozens of parameter data such as effective object number, circumference, diameter, roundness, area, content density, effective object number, effective object area, effective object area percentage, effective object density, etc. It can also output and import dataEXCELTable, number the particles captured by the camera corresponding to the particle numbers in the table.
3.The software can classify particles from small to large and label the number, area, and percentage of particles in each classification segment.
4.software automatic generationEXCELTable test report for saving and future use.
■ Instrument parameters:
1、 Polarizing microscope
1.eyepiece
Wide field eyepiece10×/22mm
Grid eyepiece10×/20mm
Cross shaped eyepiece10×/20mm
Dividing eyepiece10×/20mm
2.Infinite flat field achromatic stress free objective lens
Magnification factor:4X、10X、20X、40X、60X
Numerical aperture(NA):0.10、0.25、0.40、0.65、0.85
Working distance(mm): 37.5、7.31、8.71、0.66、0.26
3.Total magnification:40X100X200X400X600×
4.Converter: Four hole adjustable center
5.rotating platform
Platform size: Φ172mm
Mobile range:360°
6.Spotlight lens:N. A.1.30Swinging Abbe spotlight with variable light bar
7.Color filter: plug-in type color filter (green, blue, neutral)
8.Coloring device: gypsum lambda sheet, mica1/4λ plate, quartz wedge
9.Focusing: Coarse and micro coaxial focusing, using a gear rack transmission mechanism with micro grid values0.002mm
10.light source
Upper lighting: halogen lamp12V/30W,AC85V-230V,Adjustable brightness
Lower lighting: halogen lamp12V/30W,AC85V-230V,Adjustable brightness
2、 Melting point meter
1.Input power supply: AC220V ±10﹪45-60HZ
2.Hot table heating power supply: DC24V 1.5A
3.attack Consumption:75W
4.Temperature range: room temperature—400℃
5.Maximum temperature of the outer body of the hot stage: when the hot body400℃; room temperature25At ℃≤70℃
6.Maximum load capacity of the heating element:200can
7.Working mode: continuous
8.Measurement accuracy: full range≤±0.3﹪
9.System volatility:±1℃
10.Maximum heating rate(MAXLocation: Room temperature—100℃≤40second
11.Can set the slowest heating rate: 36second/℃, reaching400℃ time:4hour
12.Adjustable heating rate range:MAX—36second/℃
13.Immediate constant temperature response time:≤0.01second
14.Workbench dimensions:Φ110mmThickness:12mm
15.Workbench working face size:Φ30mmThere are two types: transmitted light and reflected light
16.Weight:7.6Kg
17.Outer packaging size:360×215×365mm
■ Instrument completeness:
1.Polarizing microscope:1tower
2.Observation tube:1only
3.Infinite flat field achromatic stress free objective lens:4X10X20X40X60X: Each1only
Infinite flat field achromatic stress free metallographic objective:40X60X: Each1only
4.10XEyepiece:2only
5.10XDividing, cross, grid eyepiece: each1only
6.Gypsum lambda test piece:1piece
7.muscovite1/4λ test piece:1piece
8.Quartz wedge:1piece
9.0.01Millimeter micrometer:1a flat
10.Blue filter:1a flat
11.Adjusting handle:2only
12.Vertical lighting:1cover
13.Spare light bulb:2only
14.Polarization melting point hot stage:1tower
15.Adaptive mirror0.5X:1only
16.Digital cameraYH-3001:1only
17.Carbon black analysis software:1cover
18.MicrometerC1:1piece