The GSGC-5411 online gas chromatograph is based on gas chromatography technology, combined with a room temperature adsorption rapid flash evaporation system, to analyze C6-C12VOCs in the sample gas. The air sample is actively extracted using a diaphragm pump, and the low-carbon volatile organic compounds in the sampled volume sample are adsorbed in a composite packed adsorption tube controlled by a high-precision mass flow controller (MFC). After flash evaporation and injection, the high boiling point compounds above C12 are blown back into the gas chromatography separation system to be pre column blowback, and the target compounds in the range of C6-C12 are detected by a hydrogen flame ionization detector (FID).
1. The composite filling adsorption tube has good adsorption performance for different types of target compounds, with a large sampling volume of up to 1L or more;
2. Rapid flash evaporation technology, with a high heating rate of up to 40 ° C/s in the adsorption tube, ensures a narrow injection band width;
3. Adopting pre column/back blowing technology to retain target components and blow back high boiling point components in the sample not only shortens the total analysis time, but also extends the service life of the analysis column and reduces operation and maintenance costs,
4. The analysis flow path adopts deactivated and passivated pipelines and devices, which have no adsorption on the sample and accurate star measurement results,
5. Use automatic electronic flow control technology (EPC) to control the carrier gas, air, and hydrogen with high precision (0.01psi), Good repeatability and reproducibility,
6. The core components are all made of well-known brands, with high reliability and long service life.
Detection capability for C6-C12 high boiling point volatile organic compounds
Daytime range of 0-300ppb adsorption tube; 0.1 ppm~1 OOppb Dingbao Ring
Detector: Hydrogen Flame Ionization Detector (FID)
Detection limit ≤ 0.03ppb (benzene)
Repeatability RSD ≤ 3% 0.03ppb (benzene)
Analysis cycle 15-30 minutes
Power supply<800W, 220V AC/50Hz
Working environment temperature (-10~50) 0c, Humidity (10%~90%) RH
The gas source requires carrier gas, high-purity nitrogen or zero grade air (≥ 99.999%); Pure hydrogen gas for combustion (≥ 99.999%); Gas assisted zero grade air (with a concentration of less than 20ppb)
Output 4-20mA RS232/RS485、 Ethernet
19 inch standard chassis, 6U
- The gas concentration inversion algorithm adopts publicly published and scientific methods
- Click to download