Product Introduction:
Instrument principle
The instrument is based on gas chromatography technology, combined with a low-temperature enrichment rapid flash evaporation system, to analyze low-carbon VOCs in the sample gas. The effective enrichment factor is more than 50 times higher than that at room temperature. The diaphragm pump extracts air samples, and the sampling volume is precisely controlled by a high-precision mass flow controller (MFC). The low-carbon volatile organic compounds in the sample are enriched in a low-temperature cold trap at minus 20 degrees Celsius, and then rapidly flashed and injected into a gas chromatography separation system with pre column blowback. High boiling point compounds above C6 are blowback, while C2-C5 low boiling point compounds are detected by a hydrogen flame ionization detector (FID).
Instrument Features
>The semiconductor refrigeration low-temperature trap can be as low as minus 20 degrees Celsius, greatly improving the enrichment efficiency of low boiling point volatile organic compounds;
>Rapid flash evaporation technology, with a maximum heating rate of up to 40 ℃/s for the enrichment tube, ensures a narrow injection band width;
>By using pre column/reverse blowing technology, the target components are retained and the high boiling point components in the sample are blown back, which not only shortens the total analysis time but also extends the service life of the analysis column, reducing operation and maintenance costs;
>The analysis flow path adopts deactivated and passivated pipelines and devices, which have no adsorption on the sample and accurate measurement results;
>Using automatic electronic flow control technology (EPC) to control carrier gas, air, and hydrogen, with high precision (0.01psi), good repeatability and reproducibility;
>The core components are all made of internationally renowned brands, with high reliability and long service life.
Application field
>Online monitoring of urban ambient air
>Online monitoring of volatile organic compounds at discharge outlets or factory boundaries in chemical industrial parks
>Online monitoring of specific pollutants
Technical Specifications
| Project | indicator parameters |
| detection capability | C2-C5 low boiling point volatile organic compounds |
| range | 0-300ppb low-temperature cold trap; 0.1ppm~100ppm quantitative ring; optional |
| detector | Hydrogen flame ionization detector (FID) |
| detection limit | ≤ 0.05 ppb (propane) |
| repetitiveness | RSD ≤ 3% (10ppb propane) |
| Analysis cycle | 15~30 min |
| Power supply | <800W,220V AC/50Hz |
| working environment | Temperature: (-10~50) ℃, Humidity: (10%~90%) RH |
| Gas source requirements | Carrier gas: high-purity nitrogen or zero grade air (≥ 99.999%); Combustion gas: high-purity hydrogen gas (≥ 99.999%) Auxiliary gas: zero grade air (hydrocarbons<20PPb) |
| output | 4~20mA、RS232/RS485、 Ethernet |
| size | 19 inch standard chassis, 6U |