1、 Brinell hardness testerXHB-3000RCparameter
|
Hardness measurement range: |
(8~650)HBW |
||
|
Brinell hardness test(N) |
include612.9N(62.5Kgf)、980N(100Kgf)、1226N(125Kgf)、1839N(187.5Kgf)、2452N(250Kgf)、4900N(500Kgf)、7355N(750Kgf)The9800N(1000Kgf)、14700N (1500Kgf)、29400N(3000kgf) |
||
|
Observation system with eyepiece, micrometer, eyepiece magnification: |
not less than20times |
||
|
Indication accuracy: |
Hardness range( HBW) ≤ 125 125<HBW≤225 > 225 |
Maximum allowable error of indication% ±3 ±2.5 ±2.0 |
Repeatability of indication% ≤ 3.5 ≤ 3.0 ≤ 2.5 |
|
Maximum allowable height of test piece(mm): |
not less than220mm |
||
|
Distance from the center of the pressure head to the body(mm): |
not less than35mm |
||
|
Minimum division value of micrometer eyepiece drum: |
not greater than0.00125mm |
||
|
Power Supply: |
AC220V/50HZ |
||
|
Dimensions of hardness tester(mm): |
760×530×185 |
||
|
Weight: |
120kg |
||
2、 Other requirements:
|
configuration |
Equipped with specialized measurement software |
|
beltUSBinterface |