BINDER VD115 vacuum drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 15 ℃ to 200 ℃
2 APT.lineTMPreheating chamber technology
3. Controller using interval programming and real-time programming
4. 2 aluminum expansion brackets
5. Inert gas interface
6. Anti impact safety glass panel with spring plates
7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
8. Computer interface: RS 422
9. Two switches output 24 V DC (maximum 0.4 A)
2、 Technical parameters:
| PRODUCT MODEL | VD115-230V1 | VD115UL-120V1 |
| Choose a model | standard | standard |
| Item code | 9030-0031 | 9030-0037 |
| measure | ||
| Internal volume [L] | 115 | 115 |
| Net weight of equipment (unloaded) [kg] | 153 | 153 |
| Maximum total load [kg] | 65 | 65 |
| Maximum load per partition [kg] | 20 | 20 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 740 | 740 |
| Net weight at height [mm] | 900 | 900 |
| Depth net weight [mm] | 670 | 670 |
| Wall to side distance [mm] | 135 | 135 |
| Wall distance from rear [mm] | 100 | 100 |
| Observation window width [mm] | 456 | 456 |
| Observation window height [mm] | 413 | 413 |
| Internal dimensions | ||
| Width [mm] | 506 | 506 |
| Height [mm] | 506 | 506 |
| Depth [mm] | 460 | 460 |
| The number of doors | ||
| Outer door | 1 | 1 |
| firmware | ||
| Number of partitions (hours/maximum) | 2/6 | 2/6 |
| Temperature performance data | ||
| The temperature range exceeds the indoor temperature by 15 ℃ to [℃] | 200 | 200 |
| Temperature deviation within 100 ℃ [± K] | 3.5 | 3.5 |
| Temperature fluctuation [± K] | 0.1 | 0.1 |
| Heating time to 100 ℃ [min] | 95 | 95 |
| Vacuum data | ||
| Allowable maximum vacuum [mbar] | 0.01 | 0.01 |
| Leakage rate [bar/h] | 0.01 | 0.01 |
| Electric data | ||
| Rated voltage [V] | 230 | 120 |
| Power frequency [Hz] | 50/60 | 60 |
| Rated power [kW] | 1.7 | 1.7 |
| Equipment safety device [A] | 10 | 20 |
| Phase (rated voltage) | 1~ | 1~ |
| Connector | ||
| Vacuum interface with small flange [DN mm] | 16 | 16 |
| Test hole with small flange [DN mm] | 16 | 16 |
| Inert gas interface with flow restrictor (RP) | 3/8 | 3/8 |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.