English

Shanghai Huanao Technology Co., Ltd
Home>Products>BINDER MKF115 High and Low Temperature Alternating Climate Chamber
BINDER MKF115 High and Low Temperature Alternating Climate Chamber
BINDER MKF115 High and Low Temperature Alternating Climate Chamber
Product details

BINDER MKF115 High and Low Temperature Alternating Climate Chamber

1、 Product Introduction:

1. Temperature range: -40 ℃ to 180 ℃

2. Humidity range: 10% relative humidity to 98% relative humidity

3. Integrated water storage tank 20 L

4. Four zero voltage switch contacts that can be activated through the MCS controller

5、APT.lineTMPreheating chamber technology

6. Programmable condensation protection for samples

7. Heated observation window with LED internal lighting

8. Using capacitive humidity sensors and steam humidification for humidity regulation

9. BINDER communication software APT-COM 3 Basic version

10. Fault diagnosis system using optical and sound alarms

11. Controller using interval programming

12. Access hole with silicone plug 50mm, left

13. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm

14. 4 casters (2 with brakes)

15. Computer interface: Ethernet

16. 230 V socket on the right side operating component

17. Display through LCD color monitor

18. Adjustable slope function function function

19. Integrated continuous automatic recorder

20. Real time clock

21. Door heating

22. Alarm message when fresh water tank is short of water

23. One socket made of stainless steel

24. Hose protection joint, including water supply and drainage pipes with a height not exceeding 1m

25. Inner chamber made of stainless steel

26. PT100 temperature sensor

27. Heat insulation through polyurethane foam

28. Fluorine free refrigerant R-404A

29. Cooling through a compression cooler

30. Finned evaporator

2、 Technical parameters:

Product Model MKF115-400V1

MKF115-400V-C1

Choose a model standard Equipped with transformer and frequency converter
Item code 9020-0283 9020-0291
measurement
Internal volume [L] 115 115
Net weight of equipment (unloaded) [kg] 280 280
Maximum total load [kg] 60 60
Maximum load per partition [kg] 30 30
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 980 980
Net weight at height [mm] 1725 1725
Depth net weight [mm] 865 865
Wall distance from rear [mm] 210 210
Wall to side distance [mm] 300 300
Observation window width [mm] 288 288
Observation window height [mm] 222 222
internal dimensions
Width [mm] 600 600
Height [mm] 480 480
Depth [mm] 400 400
number of doors
outer gate 1 1
Data related to the environment
Sound pressure level [dB (A)] 62 67
firmware
Number of partitions (hours/maximum) 1/4 1/4
Temperature performance data
Temperature range [℃] -40…180 -40…180
The temperature deviation depends on the set value [± K] 0.1…1.3 0.1…1.3
The temperature fluctuation depends on the set value [± K] 0.1…0.6 0.1…0.6
The average heating rate is based on IEC 60068-3-5 [K/min] 5.5 5.5
Cooling time from 180 ℃ to -40 ℃ [min] 120 120
The average cooling rate is based on IEC 60068-3-5 [K/min] 4.5 4.5
The maximum thermal compensation is at 25 ℃ and 90% relative humidity [W] 400 400
The maximum thermal compensation is at 25 ℃ [W] 2500 2500
Climate performance data
Temperature range [℃] 10…95 10…95
The temperature fluctuation depends on the set value [± K] 0.1…1.3 0.1…1.3
Humidity range [% relative humidity] 10…98 10…98
The humidity fluctuation depends on the set value ≤ 2.5 ± relative humidity ≤ 2.5 ± relative humidity
Condensation temperature range of condensate water [℃] 5…94 5…94
Electrical data
Rated voltage [V] 400 480
Power frequency [Hz] 50 60
Rated power [kW] 4.8 4.8
Equipment safety device [A] 16 16
Phase (rated voltage) 3~ 3~

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!