BINDER FP115 drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃
2 APT.line ™ Preheating chamber technology
3. Adjustable fan speed
4. Adjustable exhaust valve
5. Controller using interval programming and real-time programming
6. 2 chrome plated sockets
7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
8. Computer interface: RS 422
2、 Technical parameters:
| PRODUCT MODEL | FP115-230V1 | FP115UL-120V1 |
| Choose a model | standard | standard |
| Item code | 9010-0255 | 9010-0262 |
| measure | ||
| Internal volume [L] | 115 | 115 |
| Net weight of equipment (unloaded) [kg] | 62 | 62 |
| Maximum total load [kg] | 50 | 50 |
| Maximum load per partition [kg] | 20 | 20 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 835 | 835 |
| Net weight at height [mm] | 705 | 705 |
| Depth net weight [mm] | 645 | 645 |
| Wall distance from rear [mm] | 160 | 160 |
| Wall to side distance [mm] | 100 | 100 |
| Internal dimensions | ||
| Width [mm] | 600 | 600 |
| Height [mm] | 480 | 480 |
| Depth [mm] | 410 | 410 |
| Data related to the environment | ||
| Energy consumption at 150 ℃ [Wh/h] | 544 | 544 |
| firmware | ||
| Number of partitions (hours/maximum) | 2/6 | 2/6 |
| Temperature performance data | ||
| The temperature range exceeds the indoor temperature by 5 ℃ to [℃] | 300 | 300 |
| Temperature deviation is 150 ℃ [± K] | 1.8 | 1.8 |
| Temperature fluctuation at 150 ℃ [± K] | 0.3 | 0.3 |
| Heating time to 150 ℃ [min] | 30 | 30 |
| The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 8 | 8 |
| Ventilation data | ||
| Air exchange rate at 150 ℃ [x/h] | 32 | 32 |
| Electric data | ||
| Rated voltage [V] | 230 | 120 |
| Power frequency [Hz] | 50/60 | 60 |
| Rated power [kW] | 1.6 | 1.6 |
| Equipment safety device [A] | 10 | 16 |
| Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.