1、 Product Introduction:
1. Temperature range: ambient temperature plus 10 ℃ to 300 ℃
2. Up to 30% more energy-efficient than common devices on the market
3. Utilizing APT.line ™ Technology achieves high temperature accuracy
4. Adjustable fan speed
5. Controller with LCD display and advanced time function
6. Exhaust valve electromechanical control
7. 2 chrome plated sockets
8. Stackable
9. Integrated independently adjustable temperature safety device level 2 (DIN 12880) with optical alarm
10. Ergonomic handle design
11. Ethernet interface
12. USB interface for recording data
2、 Technical parameters:
| Product Model | FED056-230V1 | FED056UL-120V1 |
| Choose a model | standard | standard |
| Item code | 9010-0295 | 9010-0296 |
| measurement | ||
| Internal volume [L] | 60 | 60 |
| Net weight of equipment (unloaded) [kg] | 39 | 39 |
| Maximum total load [kg] | 70 | 70 |
| Maximum load per partition [kg] | 30 | 30 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 560 | 560 |
| Net weight at height [mm] | 625 | 625 |
| Depth net weight [mm] | 565 | 565 |
| Wall distance from rear [mm] | 160 | 160 |
| Wall to side distance [mm] | 100 | 100 |
| internal dimensions | ||
| Width [mm] | 400 | 400 |
| Height [mm] | 440 | 440 |
| Depth [mm] | 345 | 345 |
| number of doors | ||
| outer gate | 1 | 1 |
| Data related to the environment | ||
| Energy consumption at 150 ℃ [Wh/h] | 290 | 290 |
| Sound pressure level [dB (A)] | 43 | 43 |
| firmware | ||
| Number of partitions (hours/maximum) | 2/4 | 2/4 |
| Temperature performance data | ||
| The temperature range exceeds the indoor temperature by 10 ℃ to [℃] | 300 | 300 |
| Temperature deviation is 150 ℃ [± K] | 1.4 | 1.4 |
| Temperature fluctuation at 150 ℃ [± K] | 0.3 | 0.3 |
| Heating time to 150 ℃ [min] | 15 | 15 |
| The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 4 | 4 |
| Electrical data | ||
| Rated voltage [V] | 230 | 120 |
| Power frequency [Hz] | 50/60 | 60 |
| Rated power [kW] | 1.1 | 1.2 |
| Equipment safety device [A] | 6.3 | 12.5 |
| Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.