The BSC-1200 Ⅱ B2 full row external biological safety cabinet (hereinafter referred to as the safety cabinet) is an air purification equipment that provides a local dust-free and sterile working environment, and can artificially control the discharge of contaminated air in the work area through a dedicated filtering channel to avoid harm to people and the environment. It is a safe microbial specific clean workbench. It can also be applied to related industries such as biological laboratories, medical and health care, biopharmaceuticals, etc., with good effects on improving process conditions and protecting the health of operators.
|
|
BSC-1200 Ⅱ B2 full row external biological safety cabinet is a vertical unidirectional flow type local air purification equipment. BSC-1200ⅡB2Full row typePart of the air in the working area of the external biological safety cabinet is sent into the high-efficiency air filter by the top air supply fan for filtration, and then blown out from the air outlet surface to form a clean airflow. Clean airflow flows through the work area at a certain cross-sectional wind speed, thus forming a clean working environment. Another part of the air in the workspace is replenished by the intake of indoor air from the front side of the countertop. The air in the work area is drawn into the static pressure box by the fan through the return air vents on both sides of the table, and a part is filtered through the exhaust filter and discharged from the safety cabinet through the top exhaust valve. The working area of the safety cabinet is completely surrounded by negative pressure zones, effectively sealing off unclean aerosols in the working area to ensure the safety of operators.
|
|
|
Press the "Fan" button to turn on the fan, the fan indicator light will turn on, and the display window will show the numerical value. The computer simultaneously accumulates the working time of the fan and archives it.
safetyBefore leaving the factory, the air supply and exhaust volumes of the cabinet are adjusted to meet the requirements. When the fan is turned on, the air supply and exhaust volumes are automatically and constantly adjusted without the need for manual adjustment. Press the "Lighting" button to turn on the lighting. At the same time, the lighting indicator light is on. Press the "Sterilization" button to turn on the sterilization UV lamp, and at the same time, the sterilization indicator light will light up. The computer will also accumulate the working time of the UV lamp and save it.
|
Second generation control interface, easy to operate
Steel plate baking paint process, optional all stainless steel material
Sliding door limit alarm, can constantly remind the operator
Equipped with dedicated waterproof sockets and drainage interfaces, providing great convenience for operators
Convex wind wall forms a completely isolated state between the work area and the operators
The exhaust system is equipped with a dedicated high-efficiency filter to ensure that the discharged air meets national hygiene requirements
Adopting a structure where all positive pressure contaminated areas are surrounded by negative pressure zones
100% external exhaust
| texture of material:Outer steel plate painted with paint and inner stainless steel (optional all stainless steel) |
|
Cleanliness:Class A |
|
|
|
|
| Wind speed:0.25~0.45m/s, adjustable |
|
Colony count:≤ 0.5 per dish per hour (in a 90mm culture dish) |
|
|
|
|
| Front inhalation wind speed:≥0.5 m/s |
|
Box airtightness:Pressure drop ≤ 10% (at 500pa pressure for 30 minutes) |
|
|
|
|
| Half peak vibration value:≤5μm |
|
Workplace size:1200*600*650mm External dimensions:1400*800*2230mm |