Y09-5100 laser dust particles
Y09-5100 laser dust particles
Product details

Y09-5100 laser dust particle counter

characteristic:

1. Sampling volume of 100 liters per minute.

2. Dual use of AC and DC.3. Touch screen control operation.

4. Six channel particle size display, showing particle concentration (particles/m)3).

5. Automatic UCL calculation (applicable to 2010 version GMP, 1998 version GMP, ISO14644, etc.).

The Y09-5100 laser dust particle counter (100L/min) is used to measure the size and number of dust particles per unit volume of air in clean environments, and can directly detect clean environments with cleanliness levels of Class A and Class B.

The Y09-5100 laser dust particle counter adopts a semiconductor laser light source, a large touch screen display, small size, light weight, high detection accuracy, simple and clear functional operation, microprocessor control, and can store, print and measure

As a result, testing in a clean environment is very convenient. In addition, data can be downloaded and stored through specialized computer software. The product is widely used in pharmaceuticals, optics, chemistry, food, cosmetics, electronic health, biological products, aerospace, etc

Waiting for departments

Y09-5100 laser dust particle counterTechnical parameters:

Exterior Dim

268 * 192 * 292 (Wide * High * Deep)

Display Method

Touch screen display (LCD)

Weight

7.5 Kg

Maximum power consumption. Consumption

80W

Power Source

DC16.8V Built in Lithium Battery

Size Channels

0.5、1、3、5、10、25μm

Calculate Calculation

95% UCL calculation, can directly display particle concentration (particles/cubic meter, particles/cubic foot)

Sampling flow rate

100L/min Imported Pump

Test cycle Sample Time

Selectable between 0 and 999999 seconds

Self purification time Zero Count

≤10min

Measurement Location Labels

0~9999

Print Printer

Built in Thermal Printer

Alarm level

ISO/Self designed (Class)

Data Storage Date Storage

20000 sets/300 sets of UCL (with PC specific communication software) 20000 samples/300 samples of UCL (with PC software)

Data Output Date Output

USB port output from USB flash drive

Light Source

Semiconductor laser (lifespan greater than 30000 hours) Laser diode (30000 hours mean time between failures)

Maximum Rated Particle Concentration

35000 particles/L

Continuous working time Operating time

3 hours

Environmental conditions for use

Temperature: 10~35 ℃ Humidity: 20%~75% RH Atmospheric Pressure: 86~106Kpa