Adopting standards:
GB/T 212-2008 "Industrial Analysis Methods for Coal"
GB/T 483-2007 General Provisions for Coal Analysis Test Methods
GB/T 2001-1991 "Methods for analysis and determination of coke industry"
AmericaASTM D5142-2009 standard
Scope of application:
Suitable for industries such as power, metallurgy, petrochemicals, coalification, environmental protection, geological exploration, grain, and research institutions to determine the moisture, ash content, volatile matter, and fixed carbon of coal, coke, ore, and other substances.
Functional features:
●WJGF-8000APerformance characteristics of high-precision fully automatic industrial analyzer:
Highly automated: After placing the sample, simply click the mouse to automatically complete the test and print the results, with simple operation.
Accurate testing: After testing more than 20 types of standard coal samples, the results are accurate and correct. Compared with the national standard method, the testing accuracy meets the requirements.
■ High efficiency: Quick method:19The moisture, ash content, and volatile matter of each sample are ≤180min
The three indicators of moisture, ash content, and volatile matter can be measured in any combination or separately.
Adopting an imported integral furnace, it has fast heating speed, energy saving, and short testing time.
Adopting imported electronic scales and stepper motors to ensure stable testing performance.
The newly added experiment will be temporarily suspended midway, and the functionality of the experiment can be resumed the next day. New electric push rod automatic up and down switch cover. Added wireless data sharing function, allowing users to wirelessly transmit data in a targeted manner.
Can be connected to an electronic balance.
●WJGF-8000ATechnical parameters of high-precision fully automatic industrial analyzer:
■ Power: ≤5kw
Gas requirements:
Oxygen: Purity99.5%Pressure after decompression0.1Mpa
Nitrogen: Purity99.5%Pressure after decompression0.1Mpa
Pressure reducer: high-end0~25Mpalow-end0~0.4Mpa
■ Number of crucibles: Can be tested at once19A sample
Sample quality: moisture and ash content samples0.6~1.1g(Adjustable quality)
■ Furnace temperature range: room temperature~1000℃
Analysis accuracy: meets the requirementsGB/T212-2008Standard requirements for industrial analysis of coal
■ Dimensions:605×605×630(mm)
■ Weight:95kg