Soil Fertilizer Trace Metal Content Analyzer -1
According to the requirements of the soil testing formula fertilization suggestion card, the main soil data to be tested is the content of the following substances detected by this instrument:
Potassium, iron, manganese, copper, zinc, molybdenum, calcium, magnesium, nickel, and lead (with the addition of graphite furnace, silicon and boron can also be measured)
Detection of nitrogen, phosphorus, organic matter, and sulfur content using a UV spectrophotometer
Use an acidity meter to measure the pH value
This method is the flame method, which is a relatively affordable method suitable for enterprise procurement
Introduction to fully automatic flame atomic absorption spectrophotometer: It is a precision instrument mainly used to detect the content of trace metal elements in various soil and fertilizer substances. It can detect the following elements: manganese, iron, copper, zinc, lithium, sodium, magnesium, iron, potassium, calcium, chromium (g è), cobalt, nickel, arsenic, selenium (x ī), silver, cadmium, indium, tin, antimony, platinum, gold, lead and more than 40 other elements. It is mainly detected by flame method, and the detection limit can generally be PPM level/parts per million (with a graphite furnace atomizer interface, which can be added later to make the detection limit reach PPB level or one billionth content). The computer fully controls and processes the data automatically. Automatic switching turret with eight lights, spectral bandwidth, five automatic options of 0.1, 0.2, 0.4, 1.0, and 2.0nm, automatic flow setting, automatic ignition, and automatic shutdown protection; Automatically adjust negative high voltage and lamp current; Automatic setting of optimal flame height and atomizer position, automatic control of wavelength scanning, automatic peak searching, international level full band fast scanning and positioning, automatic setting of gas flow rate, automatic selection of optimal fuel assist ratio for element analysis, real-time monitoring and measurement methods for cooling water flow, flame and abnormal pressure include air acetylene flame method, graphite furnace method, hydride generator atomic absorption method, oxygen enriched air acetylene flame method; A hydride generator can be added to expand the detection range.
optical system
Wavelength range 190-900nm
Monochromator aberration correction C-T type monochromator
Dispersion element planar diffraction grating, 1800 lines/mm, engraving area 40x40mm2
Spectral bandwidths of 0.1, 0.2, 0.4, 1.0, and 2.0nm are automatically selectable in five levels
Wavelength accuracy is better than ± 0.20nm across the entire wavelength range
Wavelength repeatability across the entire wavelength range<0.10nm
Baseline drift 0.004A/30min
Detector high sensitivity, wide spectral range photomultiplier tube
Resolution: Spectral bandwidth of 0.2nm with split manganese double lines (279.5nm and 279.8nm) and valley to peak energy ratio<20%
Flame atomization system
Characteristic concentration (Cu) 0.02ug/ml/1%
Detection limit (Cu) 0.005ug/ml
100mm metal titanium burner, air-cooled premixed type
Precision RSD ≤ 0.8%
Spray High efficiency glass atomizer
雾化室 耐腐蚀材料雾化室
Position adjustment flame burner optimal height and automatic setting of front and rear positions
The security measures have multiple automatic safety protection functions
instrument parameters
host
Power supply AC220V/50Hz
Power 220VA
Weight 100kg
Volume 1170mm (length) × 480mm (width) × 440mm (height)
Standard Configuration
Host wooden box packaging and transportation, low-noise oil-free air compressor installation and commissioning, three hollow cathode lamp dedicated pressure gauges and random accessories, on-site training for users on usage methods