brand | Other brands | Price range | discuss personally |
---|---|---|---|
measuring method | Light scattering method | Measurement parameters | Other parameter combinations |
Origin category | Import | application area | Environmental protection, biotechnology, petroleum, mining, energy |
Thermo's PDR-1500 from the United StatesPM2.5Aerosol particle detector
The use of spectrophotometric turbidity method to determine particulate matter concentration can detect PM10, PM4.0, PM2.5 or PM1.0. It has the characteristics of simple operation, easy portability, and long continuous working time.
Product Introduction
Thermo's PDR-1500PM2.5 from the United StatesThe aerosol particle detector is a portable, active sampling type individual atmospheric particle detector. It uses NIOSH's 0500 and 0600 methods to measure aerosol particles and can detect dust concentrations ranging from 0.1-10 μ m to 0-400mg/m3. It is equipped with a cyclone cutter that can be traced back to ACGIH and can detect PM10, PM4, PM2.5, or PM1. It displays instantaneous values, large values, STEL values, and TWA values. It can store more than 99 sets of 500000 data
THERMO PDR-1500 Portable Aerosol Particulate Product Details from the United States
The PDR-1500 aerosol particle detector is a highly integrated portable, active sampling type individual atmospheric particle detector, characterized by high detection accuracy, small size, light weight, easy operation, and long outdoor operation time
Using spectrophotometric turbidity method to determine the concentration of particulate matter in the atmosphere; Real volume flow control with relative humidity compensation function, enabling real-time and accurate measurement of particulate matter concentration; The integrated sample filter cartridge facilitates user collection of particulate matter for data validation using weighing methods
Can detect particles ranging from 0.1-10 μ m, with a dust concentration range of 0-400mg/m3
Equipped with 2 types of cyclone cutters traceable to ACGIH (options), the predetermined pump flow rate can be set to detect PM10, PM4.0, PM2.5, or PM1.0; The spiral shaped sample inlet can ensure the representativeness of the inhaled particulate matter samples even without a cyclone cutter
Cyclone cutter (option): GK2.05 for PM-4 to PM-10, BGI SCC1.062 for PM-1 to PM-4
Real time display of instantaneous concentration, maximum value, STEL value, and TWA value
Measure aerosol particles using NIOSH's 0500 and 0600 methods; 0.8um filter (option)
The built-in memory can store over 500000 pieces of data from 99 independent data sets
● Flexible program to record data logs - time/date, total number of data points, temperature, relative humidity, atmospheric pressure, average concentration
Degree
Excellent adaptability to environmental changes; Suitable for on-site FA execution, particle size assessment, hazard assessment, quality verification, and individual testing
Exposure of data and protection, etc
Technical specifications of Thermo's PDR-1500 aerosol particle detector in the United States
Display LCD digital display; 2 lines x 16 characters
Sampling method: Built in pump suction active sampling
Measurement methods: spectrophotometric turbidity method, laser measurement
Measurement range 0.001-400mg/m3; Autoshift
Scattering coefficient 1.5 × 10-6 to 0.6m-1 @ λ=880nm
Resolution: 0.1% reading or 0.001mg/m3, whichever is higher
Accuracy ± 5% reading, traceable to SAE testing
Display update for 1 second
Reproducibility ± 2% reading or ± 0.005mg/m3, whichever is greater (1s average)
± 0.5% reading or ± 0.0015mg/m3, whichever is greater (10s average)
± 0.2% reading or ± 0.0005mg/m3, whichever is greater (60s average)
Flow range 1.0-3.5L/min, adjustable
Response particle size 0.1-10 μ m
Particle size separation of 1.0-10 μ m; Aerodynamic particle size separation method
The average time for displaying concentration data with a sampling interval of 1-60 seconds is adjustable
Alarm interval from 1 second to 1 hour, adjustable
Data storage>500000
99 sets of data
Storage format: time/date, total number of data points, temperature, relative humidity, atmospheric pressure, average concentration
Communication interface USB or RS23219200Baud
Analog output 0-5V.DC, 4-20mA; Real time output, must be within the full range of 0-0.1/0-0.4/0-1.0/0-10/0-40/0-100/0-
400mg/m3 room setting
Working environment -10-50 ℃; 10-90%RH, No condensation; Storage environment -20-70 ℃
Working power supply: 4 x AA alkaline batteries; Running time:>24 hours @ 1.2L/min,>6 hours @ 3.5L/min
Dimensions 181x143x84mm (H × W × D)
Weight 1.2kg
Standard configuration includes host, filter, battery, power adapter, RS232 interface cable, USB interface cable, communication software
Operation manual and related materials