VIP member
P270 High Performance Liquid Chromatography System
Product Introduction: Liquid chromatography technology is increasingly widely used for sample separation and purification in fields such as pharmaceut
Product details
P270Recommended configuration
|
Serial Number
|
product name
|
specifications
|
quantity
|
unit
|
|
1
|
P270 high-pressure infusion pump
|
Maximum flow rate of 99.99 ml/min
|
2
|
tower
|
|
2
|
UV230+UV Visible Detector (Preparation Type)
|
190-720nm, deuterium lamp+tungsten lamp
|
1
|
tower
|
|
3
|
Rehodyne 3725i-038 Manual Injection Valve
|
Standard configuration of 10ml quantitative ring (maximum configuration of 20ml quantitative ring)
|
1
|
branch
|
|
4
|
EC2000 Chromatographic Data Processing Workstation
|
Full Chinese interface, compliant with GLP requirements
|
1
|
cover
|
|
5
|
P270 Gradient System Startup Package
|
System Startup Package
|
1
|
cover
|
|
6
|
Preparation mixer
|
Static Mixer
|
1
|
tower
|
|
7
|
Semi prepared column
|
According to customer requirements, inner diameters of 50mm, 40mm, 30mm, 20mm, and 10mm are all acceptable
|
1
|
branch
|
|
8
|
fraction collector
|
According to customer requirements, choose between peak collection or time-based collection
|
1
|
tower
|
Note: The table shows the configuration of the P270 semi prepared gradient system, and there is also an optional P270 semi prepared isometric system.
performance index
|
P270
High pressure constant flow pump |
Flow range
|
1.00mL/min-99.99 mL/min Set step size: 0.01 mL/min
|
|
Traffic accuracy
|
≤± 1.5% (30.00mL/min, 8.5MPa, water, room temperature)
|
|
|
Traffic stability
|
RSD≤1% 30.00mL/min, 8.5MPa, Water, at room temperature
|
|
|
maximum working pressure
|
30MPa (1.00mL/min-20.00mL/min)
20MPa (20.01mL/min-99.99mL/min) |
|
|
Display pressure error
|
≤±3%
|
|
|
pressure pulsation
|
≤3.0%
|
|
|
power supply
|
AC 110V/220V,50Hz/60Hz
|
|
|
power consumption
|
260W
|
|
|
Dimensions (length x width x height)
|
420mm×280mm×175mm
|
|
|
UV230+
UV Visible Detector |
wavelength range
|
190-720nm
|
|
Wavelength Repeatability
|
±0.1nm
|
|
|
Wavelength accuracy
|
±0.5nm
|
|
|
band width
|
6nm
|
|
|
response time
|
0.0-9.9s
|
|
|
linear range
|
≥1.5AU(5%)
|
|
|
baseline noise
|
≤±1.5×10-5AU
|
|
|
baseline drift
|
≤3×10-4AU/h
|
|
|
minimum detectable concentration
|
1.0×10-8G/mL (methanol solution of naphthalene)
|
|
|
CELL
|
Optical path length: 3mm, volume: 5 μ L
|
|
|
power supply
|
AC 110V/220V,50Hz/60Hz
|
|
|
power consumption
|
110W
|
|
|
External dimensions
(Length x Width x Height) |
420mm×280mm×175mm
|
|
|
EC2000
Data processing workstation |
Communication method
|
RS232
|
|
measuring range
|
-100mV~+2V
|
|
|
Signal resolution
|
2μV
|
|
|
Area resolution
|
0.1μV·s
|
|
|
Number of channels
|
Single channel or dual channel
|
|
|
repeatability
|
≤0.01%RSD
|
Online inquiry