Ø main features
l 1-6Channels can be selected to save costs.
l High measurement accuracy and fast response speed.
l Regular automatic calibration is possible, with minimal maintenance workload.
l Color LCD displays real-time curves for easy analysis of working conditions.
l Save one month of historical data for easy recall.
l Monochromatic cold light source, with long service life and good stability.
l Multi channel high-precision programmable current signal output, suitable for subsequent automatic dosing or data acquisition systems.
Ø Main Technical Performance
1)Measurement principle: Silicon molybdenum blue light colorimetric method
2)Measurement range: (0-100)μg/L、 (0~200)μg/L、 (0-2) mg/L optional
3)Accuracy: ± 1% FS
4)Reproducibility: ± 1% FS
5)Stability: Drift ≤ ± 1% FS/24 hours
6)Response time: Initial response time of 10 minutes
7)Sampling period: about 8 minutes per channel
8)Water sample conditions: flow rate:>2 milliliters per second
Temperature: (10-45) ℃
Pressure: (10-100) KPa
9)Environmental temperature: (5-45) ℃ (accuracy decreases above 40 ℃)
10)Environmental humidity:<85% RH
11)Reagent types: 3 types
12)Reagent consumption: about 3 liters per month, a total of 3 types
13)Output signal: freely set within the range of (0-22) mA, with each channel isolated
14)Alarm signal: buzzer, relay normally open contact
15)Power supply: 220V ± 10%, 50Hz ± 1%, 50W