The Truth Optical Technology team has over 20 years of experience in particle size characterization and application development. Nanolink S900 is a new generation of high-speed nano particle size analysis system developed based on years of scientific research achievements, which is widely used in particle characterization of organic and inorganic particles, lotion, polymers, surfactants, micelles, virus antibodies, proteins and other samples.
Nanolink S900The technical characteristics of the laser nanoparticle size analysis system include:
1. Classic 90 ° dynamic light scattering technology
2. Particle size range: 0.6nm-10 μ m
3. The new generation of high-speed digital correlators has a minimum sampling time of 25ns and a dynamic range greater than 1011
4. Military pin level PMT detection unit with integrated fiber optic technology (APD optional)
5. High power solid-state laser 532nm 50mW, with automatic energy output adjustment
6. Temperature control range 0 ℃ -90 ℃, accuracy ± 0.1 ℃
7. Condensation control - dry gas blowing technology
Technical indicators:
measuring principle | dynamic light scattering |
Particle size range | 0.6nm-10μm* |
Accuracy | Better than ± 1% (average particle size, NIST traceable latex standard) |
Repeatability | Better than ± 1% (average particle size, NIST traceable latex standard) |
Minimum sample concentration | 0.1mg/ml* |
Minimum sample size | 20μl |
Measure angle | Classic 90 degrees |
Temperature control range | 0 ℃ -90 ℃ (120 ℃ optional) |
Temperature control accuracy | ±0.1℃ |
Condensation control | Dry gas blowing |
Laser | 532nm solid-state laser, maximum power 50mW, laser output automatically adjustable |
Laser safety level | Class 1 |
Correlators | Minimum sampling time 25ns, dynamic range>1011 |
detector | High sensitivity and extremely low dark current PMT (APD optional) |
Computer configuration | Intel i5 processor, 4GB memory, 250GB hard drive and monitor, mouse and keyboard |
computer interface | USB 2.0 or above |
operating system | Windows 7 or above Professional Edition |
Operating environment temperature | 5℃-40℃ |
Operating environment humidity | 10% -85% relative humidity (no condensation) |
Power requirements | AC 220V, 50/60Hz, standard grounding |
weight | 25kg |
size | 578mm x 397mm x 248mm(LxWxH) |
Note: * depends on the sample