Beijing Zhongheng Rixin Technology Co., Ltd
Home>Products>Handheld5016 Six Channel Particle Counter (Lighthaus)
Handheld5016 Six Channel Particle Counter (Lighthaus)
Handheld5016 Six Channel Particle Counter (Lighthaus) 0.5-25.0 μ m Channel Size Standards: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μ m Free Options: 0.5, 1.0, 3
Product details
detailed description:

Handheld5016 Six Channel Particle Counter (Lighthaus)Product Introduction:

638301816842144714972.jpg

0.5 micron sensitivity

0.1 cubic feet per minute (2.83 liters per minute) flow rate

6 synchronized visual monitoring of particles

0.2 to 25 microns

Compliant with JIS standards

3.8-inch (9.25cm) touch screen display interface

Detachable rechargeable lithium battery

Large capacity data storage function (3000 samples)

Concentration -4000000/cubic foot

Temperature and humidity probe

Standardized Count Data Report, FS-209E, ISO-14644-1 or EC GMP

Display interface zoom function

Built in warning system

user-friendly design

Lightweight appearance

Computer Easy Data Download


Handheld5016 Six Channel Particle Counter (Lighthaus)Technical parameters:


0.5-25.0 microns

channel size

Standard: 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 micrometers

Free options: 0.5, 1.0, 3.0, 5.0, 10.0, 25.0 microns

Counting efficiency

50% @ 0.2 microns; 100% particles>0.3 microns (per JIS)

velocity of flow

0.1 cubic feet per minute (2.83 liters per minute)

Laser source

Laser diode

Zero count degree

<1 count/5 minutes (per JIS)

calibration

MIST tracking

count mode

Automatic, manual, buzzing, concentration, cumulative/independent

Data Storage

3000 sets of sample records, cyclically cached (including particle and environmental state data, plus area time, etc.)

communication mode

RS-232 from RJ-45 to computer or printer

Software support

LMS XChange, LMS Express, LMS Express RT and LMS Express RT+, LMS Net

Environmental measurement sensor

Temperature and humidity probe: 32-122 degrees Fahrenheit (0-50 degrees Celsius) ± 1 degree Fahrenheit (0.5 degrees Celsius), 15-90% ± 2% RH

Concentration limit

4000000 pieces/cubic foot @ 5% overlap loss

touch panel

3.8 inches (9.65 cm), monochrome, 320 × 240

report

FS-209E, ISO-14644-1 & EC GMP

call the police

Built in, adjustable alarm beep, counting, low battery, sensor fault alarm

sample input

Waiting for dynamic sample probe

Sample Output

Built in high-efficiency filter (>99.997% @ 0.3 microns)

vacuum source

Built in vacuum pump, automatic flow rate control

casing

High pressure resistance, injection molded plastic

power supply

100-240 volts, 50-60 hertz

battery

Detachable rechargeable lithium battery

size

8.75 inches (length) x 5 inches (width) x 2.5 inches (height) [22.23 x 12.7 x 6.35 cm]

weight

2.2 pounds (1 kilogram)

Environmental requirements

operate

Temperature 50 ° F to 104 ° F (10 ° C to 40 ° C)/Humidity 20% to 95% No condensation

storage

Temperature 14 ° F to 122 ° F (-10 ° C to 50 ° C)/humidity 98% non condensing

accessory

Random accessories

CD operation manual, removable battery, dynamic sample probe (direct type), purification filter, temperature and humidity probe, etc

Head, power regulator, power cord, LMS Xchange data transmission software.

Alternative options

Portable outdoor compartment, equipped with a dynamic sample probe (with a 6-foot pipeline to extend the input of dynamic samples), detachable

/Rechargeable lithium battery External battery charging tile/AC power and power adapter, thermal printer and connection

Line, LMSExpress software, LMSExpress RT software, LMS Express RT+software, micro multi-channel

Point by point particle counting technology, certification documents




Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!