Provide professional magnetic field measurement solutions | ||||||
Strong magnetic field measuring instrument(0 - 20T) | ||||||
model |
principle |
type |
number of axles |
DCaccuracy |
maximum range |
Highest resolution |
X113 |
hall |
hall sensor |
1 |
<> |
15T |
TC: 30ppm/K |
G80 |
hall |
handheld |
1 |
2% |
2T |
10μT |
G81 |
hall |
handheld |
1 |
0.8% |
10T |
1μT |
G82 |
hall |
handheld-high frequency |
1 |
0.8% |
10T |
1μT |
G83 |
hall |
handheld |
3 |
0.8% |
10T |
1μT |
G92 |
hall |
handheld |
1 |
1% |
10T |
10μT |
G93 |
hall |
handheld |
3 |
1%, 0.5% |
20T |
1μT |
G201 |
hall |
Desktop computer |
1 |
0.2% |
10T |
1μT |
G203 |
hall/reluctance |
Desktop computer |
3 |
0.2% |
10T |
10nT |
G401 |
hall |
Desktop computer |
1 |
0.04% |
10T |
1μT |
G403 |
hall |
Desktop computer |
3 |
0.04% |
10T |
1μT |
GSP301A |
hall |
Analog transmitter |
1 |
1%, 0.2% |
10T |
Output: ±3V |
GSP301D |
hall |
Digital transmitter |
1 |
1%, 0.2% |
10T |
1μT |
GSP303D |
hall |
Digital transmitter |
3 |
1%, 0.2% |
20T |
1μT |
GSP303D-S |
hall |
Digital transmitter |
3 |
0.5%, 0.1% |
3T |
1μT |
Magnetic field testing system | ||||||
GAS3000 |
Array magnetic field detection system, synchronous detection up to128The magnetic field at a certain location |
|||||
GMS980 |
High precision Gaussian meter calibration system (High precision Tesla meter calibration device) |
|||||
Weak magnetic field measuring instrument(0- 2mT) | ||||||
Weak magnetic field Gaussian meter | ||||||
model |
principle |
type |
number of axles |
DCaccuracy |
maximum range |
Highest resolution |
GMR61 |
reluctance |
handheld |
1 |
0.8% |
600μT |
10nT |
GMR63 |
reluctance |
handheld |
3 |
0.8% |
600μT |
10nT |
GF601 |
fluxgate |
handheld |
1 |
0.5% |
1mT |
0.1nT |
GF603 |
fluxgate |
handheld |
3 |
0.25% |
1mT |
0.1nT |
GF633 |
fluxgate |
Desktop computer |
3 |
0.1% |
100μT |
0.01nT |
GFP703 |
fluxgate |
Smart Transmitters |
3 |
0.5% |
1mT |
0.1nT |
GFP703S |
fluxgate |
Spectrum analyzer |
3 |
0.5% |
1mT |
0.01nT |
GFP803 |
fluxgate |
Smart Transmitters |
3 |
0.2% |
1mT |
0.1nT |
Communication magnetic field measurement(DC- 1MHz) | ||||||
model |
principle |
type |
number of axles |
ACaccuracy |
maximum range |
frequency response |
AMS-2K |
alternating current coil |
Analog sensor |
1 |
1% |
3mT |
30Hz- 2kHz |
AMS-1M |
alternating current coil |
Analog sensor |
1 |
1% |
3mT |
2kHz- 1MHz |
G1000 |
reluctance |
Smart Transmitters |
3 |
DC- 1% AC- 2% |
400μT/ Axis |
DC- 1MHz |
GA1000 |
reluctance |
handheld |
3 |
2% |
400μT/ Axis |
30Hz- 1MHz |
GSP301HFA |
reluctance |
Analog transmitter |
1 |
1% |
150μT |
20Hz-1.2MHz |
GSP303HFA |
reluctance |
Analog transmitter |
1 |
1% |
150μT |
20Hz-1.2MHz |
fluxgate sensor | ||||||
model |
principle |
type |
number of axles |
noise level |
maximum range |
output |
F901 |
fluxgate |
Economic type |
3 |
10 to 20pT |
1000μT |
±10V,single-ended |
F902 |
fluxgate |
low noise |
3 |
<> |
100μT |
±10V,single-ended |
F903 |
fluxgate |
low-power consumption |
3 |
10 to 20pT |
200μT |
±3V,finite difference |
F904 |
fluxgate |
low-power consumption|low noise |
3 |
<> |
100μT |
±3V,finite difference |
F905 |
fluxgate |
high-performance|low noise |
3 |
<> |
100μT |
±10V,single-ended |
F906 |
fluxgate |
low-power consumption |
3 |
10 to 20pT |
100μT |
±3V,finite difference |
F23 |
fluxgate |
split type|large range |
3 |
<> |
1500μT |
±10V,single-ended |
F53 |
fluxgate |
High temperature type |
3 |
≤ 300pT@175℃ |
100μT |
±5V,finite difference |
Magnetic flux gate sensor data acquisition unit | ||||||
FDU301 |
Connectable1A three-axis flux gate sensor |
|||||
ASA1000 |
Connectable2-126A single axis flux gate sensor or2-42A three-axis flux gate sensor |
|||||
AFS |
Array flux gate sensor detection system, capable of connecting multiple flux gate magnetic field transmitters |
|||||
GMR-16 |
Micro weak magnetic field sensor array, capable of synchronous acquisition2-16Data from a miniature weak magnetic field sensor |
|||||
GR100 |
Digital three-axis flux gate gradiometer |
|||||
remarks: 1Magnetic field unit conversion:1T= 10kG; 1mT= 10G; 1μT= 10mG; 1nT= 10μG 2Click on the left side of the table abovemodelYou can refer to the product information of the corresponding model |
F23Three dimensional flux gate sensor
sketch:
F23The 3D flux gate sensor is a magnetic field sensor with independent intellectual property rights that can be mass-produced by Koley Company after years of effort. Koley Company has the ability to design new shapes and analog output characteristics based on user needs, and can integrate acceleration sensors, temperature sensors, etc. into flux gate sensors.F23The typical application scenarios of 3D flux gate sensors include geomagnetic detection, package detection, underwater magnetic field monitoring, traffic monitoring, residual magnetism measurement, weak magnetic field measurement, laboratory, etc.
F23Three dimensional flux gate sensor with selectable range±70μT、±100μT、±500μT、±1000μTand±1500μTFrequency responseDC-3kHzThe linearity is less than0.01%The three-axis orthogonality error is less than1°And noise<20pt rms/√hz='' at=''>20pt>. Use24positionADCAfter digitization,DCResolution can reach up to0.1nT.
F23The 3D flux gate sensor has two packaging specifications:F23andF23SAmong themF23It is an integrated design that encapsulates the 3D flux gate probe and circuit module together;F23SIt is a split type, small probe (size)24x24x40mm)Independently packaged, through1.5Meter cable, connecting circuit modules, suitable for measuring in narrow spaces. Product according toMTBF(Mean time between failures)10Annual design.
characteristic | |
l Three dimensional flux gate sensor l Linearity:<> l Three axis orthogonality error:<> l Power supply:±12VDC ~ ±16VDC l analog signaloutput±10V |
l range±70μT ~ ±1500μTAny optional l Accuracy:<> l Noise level:<20pt rms/√hz='' ='' at=''>20pt> l Frequency response range:DC- 3kHz l Two packaging specifications: integrated and split |
F23Three dimensional flux gate sensor parameters:
model |
F23 / F23S |
|||||
performance parameter | ||||||
number of axles |
3 (XYZRight-Handed Coordinate System) |
|||||
range |
X, Y, Z |
±70μT |
±100μT |
±500μT |
±1000μT |
±1500μT |
vectorR |
121μT |
173μT |
866μT |
1732μT |
2598μT |
|
accuracy |
<> |
|||||
noise level |
<20pt rms/√hz='' ='' at='' 1hz =''>20pt>range≤±100μT) |
|||||
Start Time |
150ms |
|||||
Warm-up time |
15mins |
|||||
Analog Output |
±10V (single-ended output) |
|||||
sensitivity |
143mV/μT |
100mV/μT |
20mV/μT |
10mV/μT |
6.6mV/μT |
|
Absolute zero point |
< ±100nt =''>range≤±100μT) |
|||||
Typical zero drift |
±0.5nT/℃(range≤ ±100μT) |
|||||
Typical temperature coefficient |
±100ppm/℃(range≤ ±100μT) |
|||||
Three-axis orthogonality error |
< 1°=''> |
|||||
Linearity |
<>least-squares fit) |
|||||
frequency response(-3dB) |
> 3kHz |
|||||
hysteresis |
<2nt>2nt>Exposed to magnetic field environments up to twice the range) |
|||||
MTBF |
10year(Mean time between failures) |
|||||
Mechanical dimensions | ||||||
IntegratedF23 |
24 x 24 x 125mm |
|||||
split typeF23S |
Probe:24x24x40mmCircuit module:30 x 25 x 107mm; Connection cable length:1.5m |
|||||
Electrical specifications | ||||||
supply voltage |
±12VDCto±16VDC |
|||||
Typical power consumption |
240mW |
|||||
Environmental requirements | ||||||
operation temperature |
- 40℃to +60℃ |
|||||
storage temperature |
- 40℃to +70℃ |
|||||
protection grade |
IP67 |
Terminal Description
Cable color |
connecting terminal |
describe |
brown |
X |
Analog signal single ended output(XAxis,±10V) |
yellow |
Y |
Analog signal single ended output(YAxis,±10V) |
green |
Z |
Analog signal single ended output(ZAxis,±10V) |
blue |
AGND |
signal ground |
black |
PGND |
Power Ground |
red |
P13 |
DC IN (±12VDCto±16VDC) |
white |
N13 |
Note: Sensitivity is equal to±10VDivided by the range of the flux gate sensor, please refer to the parameter table for details.
Selection of Three Dimensional Flux Gate Sensors
F23 |
-100 |
-1m |
Three dimensional flux gate sensor types F23Integrated 3D Flux Gate Sensor F23SSplit type 3D flux gate sensor
|
Sensor range -70:70μT/axle -100:100μT/axle -500:500μT/axle -1000:1000μT/axle -1500:1500μT/axle |
Cable Length -1m:1rice -2m:2rice -5m:5rice ...... |
Note:
1Model example:F23-100-5mThis is an integrated 3D flux gate sensor(F23)The range is100μT/axle(100),Cable Length5rice(5m);
2The longest cable length100m,100Mi Nei can be customized;
3Maximum sensor range1500μT,1500μTThe interior can be customized.