Detection Principle
The YC-108S economical COD rapid analyzer is designed and developed based on the authoritative certification method HJ/T399-2007 "Determination of Chemical Oxygen Demand in Water Quality - Rapid Digestion Spectrophotometer" standard. It uses a closed digestion tube for closed digestion. In a strong acidic solution, a special oxidant containing a certain amount of potassium dichromate is used to dissolve at a constant temperature of 165 ℃ under the action of a catalyst, so that reducing substances in the water are oxidized. The absorbance of Cr6+and Cr3+in the water sample is measured at different wavelengths, and the COD content in the water is calculated by the instrument using the absorbance.
Technical Parameter
Measuring Item |
Chemical Oxygen Demand (COD) |
Measurement range |
0-10000mg/L |
accuracy |
±5% |
detection limit |
5mg/L |
Optical stability |
≤0.001A/20minute |
Measurement time |
30minute16Branch sample/write instructions or comments on |
Temperature indication error |
±2℃ |
Temperature uniformity |
±1℃ |
Light source lifespan |
1010000 hours |
Number of curves |
standard3strip extend77strip |
storing data |
1800individual |
display |
LCDdisplay |
Display interface |
All in Chinese |
Colorimetric method |
Colorimetric tube |
printer |
Thermal printer |
data interface |
USB |
ambient temperature |
5-40℃ |
ambient humidity |
Relative humidity ≤85% |
safety protection |
Dual security protection |
Pre set digestion mode |
CODTotal phosphorus and total nitrogen |
Reminder alarm |
buzzer |
Customer customization |
support |
rated voltage |
AC220V±10%/50Hz |
rated power |
instrument<10W, digestion apparatus<350W |
Alarm prompt1 |
Reaching the specified temperature and maintaining a constant temperature |
Alarm prompt2 |
Dissolve completed and stop heating |
Instrument dimensions |
310mm*230*150 |
Dimensions of the digestion instrument |
230*340*130(mm) |
Instrument characteristics
1. Complies with the HJ/T399-2007 standard "Determination of Chemical Oxygen Demand in Water Quality - Rapid Digestion spectrophotometric method", with accurate and effective measurement.
2. Adopting imported high brightness and long-life cold light source, with excellent optical performance and a light source lifespan of 100000 hours.
3. Large LCD screen with full Chinese display, direct data reading, simple and time-saving operation.
4. Integrated digestion and colorimetric analysis, no need to replace the tube, simple and fast measurement, and no safety hazards.
5. It can save 80 standard curves and 1800 measured values (date, time, parameters, test data).
6. Memory standard working curve, users can also calibrate the curve as needed.
7. One click factory reset, which can quickly recover in case of curve loss caused by misoperation.
8. It has data storage function and power-off protection function, making it convenient to query historical data and prevent data loss.
Standard configuration
Serial Number |
name call |
quantity |
Serial Number |
name call |
quantity |
1 |
Economic COD RapidmeasureFixed instrument host |
1tower |
6 |
Simplified operation flowchart |
1share |
2 |
⌀16mmColorimetric tube |
10branch |
7 |
operation instructions |
1share |
3 |
CODTesting reagents |
1cover |
8 |
Color comparison rack |
1individual |
4 |
Color tube cleaning cloth |
1piece |
9 |
certificate of conformity/warranty card |
1share |